Becton Dickinson & Co.: BD MAX SARS-CoV-2 Reagents, Cat. No. 445003 (UDI 00382904450038) and 445003-01 (UDI (01)603829044500301) - Product Usage: intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.

Recall number
Z-2754-2020
Recalling firm
Becton Dickinson & Co.
Product
BD MAX SARS-CoV-2 Reagents, Cat. No. 445003 (UDI 00382904450038) and 445003-01 (UDI (01)603829044500301) - Product Usage: intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.
Status
Terminated
Initiated
2020-07-20
Posted
not on file
Terminated
2021-07-13
Reason
BD is experiencing elevated rates of potential false positive results from certain customers when using the BD SARS-CoV-2 Reagents for the BD MAX System. False positive test results could lead to a delay in diagnosis and potentially cause the patient to be exposed to COVID-19 if they are isolated with an infected individual.
Root cause
Under Investigation by firm
Action
On July 20, 2020, the firm sent Urgent Medical Device Product Advisory letters to their customers. Customers were informed that all lots are impacted by the elevated rates of potential false positive results. Customers were advised that all positive results are considered presumptive. BD recommends that testing for any positive samples be repeated using an alternate authorized test method. Customers were asked to share the notification with all users of the reagents within their facility to ensure that they are also aware of this product advisory. Customers are to complete and return the customer response form to the recalling firm. US customers who require further assistance may contact BD Customer/Technical Support at 800-638-8663, Monday-Friday, between 7AM-7PM (EST). OUS customers may contact their local BD representative or distributor.
Quantity
30,392 to date
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GU, HI, IL, IN, KY, MD, MI, MO, NC, NJ, NY, PA, SC, TN, TX, WI, WY and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Hong Kong, Japan, Mexico, New Zealand, Peru, Philippines, Singapore.
Lot, serial or model codes
All lots are affected
510(k) numbers
not on file
PMA numbers
not on file
Product code
QJR
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register