FDA Medical Device Recalls (openFDA)
Merge Healthcare, Inc.: Merge Eye Care PACS Viewer 5.2 Merge Eye Care PACS is a software solution for the display, management, archive, interface and integration of ophthalmic device reports, images and data. Merge Eye Care PACS is a software solution using databases for patient demographics, server and other systems of intercommunication with hospital systems via HL7 and DICOM to provide to clinicians a single image viewing and management solution of images imported from various contributing devices throughout the clinical environment. Merge Eye Care PACS provides the ability to review data from any network-connected computer, and is protected by appropriate security login which permits only authorized user access. Symphony uses 256 bit or greater encryption via secure socket layer (SSL) to assure a network environment which is secure.
- Recall number
- Z-2754-2017
- Recalling firm
- Merge Healthcare, Inc.
- Product
- Merge Eye Care PACS Viewer 5.2 Merge Eye Care PACS is a software solution for the display, management, archive, interface and integration of ophthalmic device reports, images and data. Merge Eye Care PACS is a software solution using databases for patient demographics, server and other systems of intercommunication with hospital systems via HL7 and DICOM to provide to clinicians a single image viewing and management solution of images imported from various contributing devices throughout the clinical environment. Merge Eye Care PACS provides the ability to review data from any network-connected computer, and is protected by appropriate security login which permits only authorized user access. Symphony uses 256 bit or greater encryption via secure socket layer (SSL) to assure a network environment which is secure.
- Status
- Terminated
- Initiated
- 2017-06-13
- Posted
- 2017-07-10
- Terminated
- 2020-09-17
- Reason
- When the user has not set up any user preference on the sorting order to render the study images (OS/OD/etc.) and reports, the ECP may get an incorrect study index. If the user attempts to delete an image, ECP may then read that image as OS when it is, in fact, OD.
- Root cause
- Software design
- Action
- The firm initiated the recall by certified letter the week of 06/26/2017. The notice provided the work-around and requested a response.
- Quantity
- 18 units
- Distribution
- FL, IL, CO, MI, CA, NY, ND, OH, CT
- Lot, serial or model codes
- Version 5.2
- 510(k) numbers
- ["K110006"]
- PMA numbers
- not on file
- Product code
- NFJ
- Firm FEI
- 1000116130
- Firm city
- Hartland
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28