Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Vascular: Medtronic 6F Zuma Guide Catheter with AL 1.5 Curve with side holes Catalog Number: Z26AL15SH The Medtronic Guiding Catheter is designed to provide a pathway through which therapeutic devices are introduced. The guiding catheter is intended to be used in the coronary or peripheral vascular system.

Recall number
Z-2754-2011
Recalling firm
Medtronic Vascular
Product
Medtronic 6F Zuma Guide Catheter with AL 1.5 Curve with side holes Catalog Number: Z26AL15SH The Medtronic Guiding Catheter is designed to provide a pathway through which therapeutic devices are introduced. The guiding catheter is intended to be used in the coronary or peripheral vascular system.
Status
Terminated
Initiated
2011-06-08
Posted
not on file
Terminated
2012-04-17
Reason
Mislabeled; Product labeled as 6F Zuma Guide with AL 1.5 curve with side holes may contain a 5F Zuma Catheter with SAL 1.0 curve
Root cause
Labeling Change Control
Action
The firm, Medtronic, sent an "Urgent Medical Device Recall Notice(for US)" letter dated June 7, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to segregate the product so that it will not be used. A Medtronic Sales Representative will visit the customers to account for all units they received and initiate a credit as appropriate. If you have any questions, please call (978) 739-0371 or (978) 739-3259.
Quantity
5 units
Distribution
Nationwide distribution: USA including states of: CA, FL, and MA.
Lot, serial or model codes
Lot # : 0005199588
510(k) numbers
["K010579","K990707"]
PMA numbers
not on file
Product code
DQY
Firm FEI
1220452
Firm city
Danvers
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register