Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter

Recall number
Z-2753-2026
Recalling firm
Boston Scientific Corporation
Product
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
Status
Open, Classified
Initiated
2026-06-11
Posted
2026-07-17
Terminated
not on file
Reason
Boston Scientific is replacing specific FARASTAR PFA generators which have a configuration that is not compatible for use with the FARAPOINT PFA catheter.
Root cause
Under Investigation by firm
Action
Boston Scientific issued an Urgent Field Safety Notice - Urgent Medical Device Recall notice to its consignees out side the US on 6/11/2026 via hand delivered letter. The notice explained the issue and requested the following actions be taken: Instructions: 2- Allow Boston Scientific representative(s) to replace affected FARASTAR PFA generators. 3- Please complete the attached Verification Form even if you do not have any product to return and send it to your local Boston Scientific office 4- If you have products to return, please package them in an appropriate shipping box. After receipt of the Verification Form, Boston Scientific will contact you to arrange return. 6- Please pass this notice to any health professional from your organization that needs to be aware, and to any organization where the potentially affected devices have been transferred (if appropriate). Please provide Boston Scientific with details of any affected devices that have been transferred to other organizations (if appropriate). The National Competent Authority has been informed of this communication.
Quantity
6 units
Distribution
International distribution in the countries of Italy, Austria, France, Germany, Belgium.
Lot, serial or model codes
GTIN/UDI 00810087180126, Serial Numbers: 1-217-0053, 1-21C-0078, 1-223-0099, 1-225-0106, 1-225-0107, 1-229-0172.
510(k) numbers
not on file
PMA numbers
["P230030","P230030S012"]
Product code
QZI
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register