Skip to the content

FDA Medical Device Recalls (openFDA)

Kapp Surgical Instrument Inc: Bariatric Target Ring (Nitinol) (a thin, circular nitinol ring) Model # K-9536-N, Lot #08-001, Lot #08-002 & Lot #08-003. One ring per plastic pouch, three rings are then placed into a plastic box with lid. This box is then placed into a solid white box. Item is shipped and sold non-sterile. It is intended to be used to radiographically (with X-Ray) mark the site of the "gastropexy" performed during gastric bypass surgery, thereby making it accessible for future diagnostic, therapeutic and feeding purposes.

Recall number
Z-2753-2011
Recalling firm
Kapp Surgical Instrument Inc
Product
Bariatric Target Ring (Nitinol) (a thin, circular nitinol ring) Model # K-9536-N, Lot #08-001, Lot #08-002 & Lot #08-003. One ring per plastic pouch, three rings are then placed into a plastic box with lid. This box is then placed into a solid white box. Item is shipped and sold non-sterile. It is intended to be used to radiographically (with X-Ray) mark the site of the "gastropexy" performed during gastric bypass surgery, thereby making it accessible for future diagnostic, therapeutic and feeding purposes.
Status
Terminated
Initiated
2011-05-24
Posted
2011-07-07
Terminated
2013-02-19
Reason
During an FDA Inspection on 3/21/11 - 3/22/11 at Kapp Surgical an uncontrolled instruction for use was discovered which contained sterilization parameters which differed from the instructions contained in the device master record. Kapp Surgical decided to conduct the recall because we cannot be sure that incorrect instructions were not shipped.
Root cause
Labeling mix-ups
Action
Kapp Surgical Instrument, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 24, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory and quarantine any product affected by this recall. The firm will make arrangements for returning the products. Replacements will be made by Kapp Surgical. Questions about this recall are directed to the QA Manager at 800-282-5277.
Quantity
222 units (rings)
Distribution
Product was distributed to the following states: OH, PA & SC. Product was also shipped to Saudia Arabia.
Lot, serial or model codes
M/odel # K-9536-N, Lot #08-001, Lot #08-002 & Lot #08-003.
510(k) numbers
["K072216"]
PMA numbers
not on file
Product code
NEU
Firm FEI
1522875
Firm city
Warrensville Heights
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register