Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Company: BD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection

Recall number
Z-2751-2017
Recalling firm
Becton Dickinson & Company
Product
BD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
Status
Terminated
Initiated
2016-09-16
Posted
not on file
Terminated
2018-03-20
Reason
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required of 10-6.
Root cause
Process control
Action
BD sent recall letter/return response form on 9/16/2016.
Quantity
648,000 units
Distribution
Nationwide Distribution
Lot, serial or model codes
6061746 2021-02 6082982 2021-02 6089821 2021-03 6117858 2021-04 6117863 2021-04 6117889 2021-04 6117891 2021-04 6146962 2021-04 6173697 2021-05 6173767 2021-06 6173779 2021-06 6208614 2021-07
510(k) numbers
["K110771"]
PMA numbers
not on file
Product code
FMF
Firm FEI
2243072
Firm city
Franklin Lakes
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register