Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958.

Recall number
Z-2746-2017
Recalling firm
Medtronic Navigation, Inc.
Product
Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958.
Status
Terminated
Initiated
2017-05-10
Posted
2017-06-26
Terminated
2021-01-08
Reason
Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated Elevate Inserter with the StealthAiR Spine Frame during spine surgical procedures.
Root cause
Software in the Use Environment
Action
Medtronic sent an Urgent - Product Advisory Notice dated May 10, 2017, to all affected customers. Instructions indicated that a Medtronic representative will be in contact to update the system with version 26 and will remove any copies of the Spine Tools Install CD, version 25. Please quarantine any copies of version 25 until they can be removed. If you have further questions regarding this communication, please contact your local Medtronic Navigation Technical Services Representative, your local Medtronic Spine Sales Representative, or call (800 595-9709.
Quantity
108
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Brazil, Germany, Switzerland and Great Britain.
Lot, serial or model codes
Kits that include affected 9731958 INSTALL CD: INST SET 9734833 UNIVERSAL NAVLOCK, INST SET 9734834 CAPSTONE TRIALS, INST SET 9734835 CLYDESDALE TRIALS, INST SET 9734836 TLIF/DLIF INSERTER, INST SET 9735283 SOLERA 5.5/6.0 DRIVERS, INST SET 9735284 SOLERA NONCAN TAPS, INST SET 9735285 SOLERA ILIAC TAPS, TAPS KIT 9735465 SOLERA AWL-TIP TAPS, REF SET 9735500 SPINE REFRENCING, UPGRD KIT 9735522 SPINE REVISION, INSTR SET 9735607 VERTEX SELECT
510(k) numbers
["K050438"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register