Maquet Cardiovascular Us Sales, Llc: The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table immediately before, during and after surgical interventions as well as for examination and treatment. The fixture is used to adapt a range of different head and horseshoe head rests with a central square mount

Recall number
Z-2746-2016
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Product
The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table immediately before, during and after surgical interventions as well as for examination and treatment. The fixture is used to adapt a range of different head and horseshoe head rests with a central square mount
Status
Terminated
Initiated
2016-08-12
Posted
not on file
Terminated
2017-05-04
Reason
Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken.
Root cause
Use error
Action
Maquet sent an "Urgent Product Recall Medical Device Correction Letter and Medical Device Correction Response Form" dated August 12, 2016 via UPS to their affected customers. Customers were advised to inspect all devices that include the fixture for damage before every use. Customers were asked to fill out and sign the enclosed Medical Device Correction Response Form and return it via email to FieldAction@maquet.com or fax to 1-973-909-9927. Customers with questions were advised to contact their MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383. For questions regarding this recall call 973-709-7487.
Quantity
19 units (US) and 1975 units (OUS)
Distribution
Nationwide Distribution to CA, FL, IA, IN, MO, UT and WI
Lot, serial or model codes
Part Number/Catalog Number 1002.64A0
510(k) numbers
not on file
PMA numbers
not on file
Product code
FQO
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register