Skip to the content

FDA Medical Device Recalls (openFDA)

Terex Utilities: Power Over Fiber Base Module

Recall number
Z-2745-2026
Recalling firm
Terex Utilities
Product
Power Over Fiber Base Module
Status
Open, Classified
Initiated
2026-05-11
Posted
2026-07-30
Terminated
not on file
Reason
Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement
Root cause
Radiation Control for Health and Safety Act
Action
A letter was received stating the product, problem and actions to be taken. The following actions being taken include: A. Develop and carry out a process to remove incorrect labeling and install all labels required under IEC 60825-1 2014. B. Generate, detail & submit full CAP (Corrective Action Plan) for FDA approval C. Generate initial laser product report or supplemental as appropriate. D. Equipment update to be performed by Terex personnel at customer location E. Procedures: * Fully certify device as Class 1 under the guidelines of FDA Laser Notice #56 by applying the IEC 60825-1 2014 laser certification standard * Update and distribute compliant UI (User Information) * Update Sales information as appropriate * Generate and apply administrative and safety labeling * Formalized quality control procedures and record keeping. If you have any questions, contact Director of Engineering at +16058824000 or email: dan.brenden@terex.com
Quantity
101
Distribution
U.S. Nationwide Distribution and International Distribution
Lot, serial or model codes
Power Over Fiber Base Module Identification P/N 633232 P/N 622365 P/N 633383 P/N 633910 P/N 635731
510(k) numbers
not on file
PMA numbers
not on file
Product code
RFN
Firm FEI
3005768141
Firm city
Watertown
Firm state
SD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register