Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature.

Recall number
Z-2742-2015
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature.
Status
Terminated
Initiated
2015-08-07
Posted
2015-09-09
Terminated
2016-05-31
Reason
Inaccuracies in SvO2, temperature, pH, pCO2, pO2, Hematocrit, and Potassium readings following a software upgrade to version 1.69.
Root cause
Software design
Action
A voluntary Urgent Medical Device Correction notice that clearly explains initial in-vivo calibration requirements, device operating ranges, and temperature measurements was sent on 08/17/2015, via express mail to consignees of CDI System 500 v1.69. Following the initial notice, Terumo will be updating the Operators Manual and will send the new manual to each consignee when available. One manual per unit at each facility will be provided. Customers with questions and return response forms may contact: Terumo Recall Email: tcvs.recall@terumomedical.com Terumo Recall Fax: 734-741-6149 Terumo CVS Customer Service: 1.800.521.2818, Monday - Friday, 8 a.m. 6 p.m. ET.
Quantity
4638
Distribution
Worldwide Distribution-US (nationwide) including DC and PR and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Australia, Belgium, Canada, Chile, Colombia, Hong Kong, Indonesia, Japan, Malaysia, Mexico, Philippines, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates.
Lot, serial or model codes
500A, 500AHCT, 500AV, 500AVHCT CDI Blood Parameter Monitoring System 500; software version 1.69.
510(k) numbers
K133658
PMA numbers
not on file
Product code
DRY
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source