Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular, LLC: The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.

Recall number
Z-2741-2026
Recalling firm
Maquet Cardiovascular, LLC
Product
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
Status
Open, Classified
Initiated
2026-06-02
Posted
2026-07-15
Terminated
not on file
Reason
Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
Root cause
Under Investigation by firm
Action
Getinge issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/18/2026 via FedEx. The notice explained the issue, risk to health, and requested the following actions be taken: " Immediately identify and quarantine all affected Vasoview Hemopro 2 devices from lot 3000547797. " Do not use affected devices. " If an affected device was used, no further intervention is required. Standard post-operative wound care per facility protocol should be followed. " If any affected devices were transferred within your organization or to another organization, ensure this notice is shared with the appropriate individuals. " Contact Getinge Customer Service 1-888-880-2874 or your Getinge representative to coordinate return of the affected units and replacement product. " Complete and return the enclosed response form to Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.
Quantity
205 units
Distribution
US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
Lot, serial or model codes
UDI 00607567700406, Serial/Lot Number 3000547797
510(k) numbers
["K101274"]
PMA numbers
not on file
Product code
GEI
Firm FEI
2242352
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register