Mako Surgical Corporation: MAKO RIO Standard System, internal cPCI Card CageP/N: 201337. Intended to assist the surgeon in providing software defined spatial boundaries during orthopedic procedures.
- Recall number
- Z-2741-2017
- Recalling firm
- Mako Surgical Corporation
- Product
- MAKO RIO Standard System, internal cPCI Card CageP/N: 201337. Intended to assist the surgeon in providing software defined spatial boundaries during orthopedic procedures.
- Status
- Terminated
- Initiated
- 2017-06-07
- Posted
- not on file
- Terminated
- 2020-06-16
- Reason
- An intermittent electrical problem that could lead to a loss of system power due to a loose screw connection.
- Root cause
- Employee error
- Action
- Customers were sent notices titled "Urgent Medical Device Correction Notification" on 06/12/2017 via UPS Next Day Mail. The firm notifies consignees of the reason for recall and instructs that users of the recalled product be provided the recall notice. Users may continue to use their MAKO system; and, Stryker/MAKO Service personnel will contact users o schedule time to inspect and replace the card cage, if necessary. Also, users are instructed to return the completed Business Reply Form within 5 business days. The completed form can be sent to strykerortho7391@stericycle.com or faxed to 855-544-4807. Any questions can be directed to Clayton Odor at 954-628-0502.
- Quantity
- 20
- Distribution
- Distributed to the states of AZ, CA, CO, CT, FL, GA, ID, LA, MA, MO, MS, NC, OH, PA, TN, TX, & WA.
- Lot, serial or model codes
- Lot No. 05212015
- 510(k) numbers
- ["K112507"]
- PMA numbers
- not on file
- Product code
- OLO
- Firm FEI
- 3005985723
- Firm city
- Davie
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28