Mako Surgical Corporation: MAKO RIO Standard System, internal cPCI Card CageP/N: 201337. Intended to assist the surgeon in providing software defined spatial boundaries during orthopedic procedures.

Recall number
Z-2741-2017
Recalling firm
Mako Surgical Corporation
Product
MAKO RIO Standard System, internal cPCI Card CageP/N: 201337. Intended to assist the surgeon in providing software defined spatial boundaries during orthopedic procedures.
Status
Terminated
Initiated
2017-06-07
Posted
not on file
Terminated
2020-06-16
Reason
An intermittent electrical problem that could lead to a loss of system power due to a loose screw connection.
Root cause
Employee error
Action
Customers were sent notices titled "Urgent Medical Device Correction Notification" on 06/12/2017 via UPS Next Day Mail. The firm notifies consignees of the reason for recall and instructs that users of the recalled product be provided the recall notice. Users may continue to use their MAKO system; and, Stryker/MAKO Service personnel will contact users o schedule time to inspect and replace the card cage, if necessary. Also, users are instructed to return the completed Business Reply Form within 5 business days. The completed form can be sent to strykerortho7391@stericycle.com or faxed to 855-544-4807. Any questions can be directed to Clayton Odor at 954-628-0502.
Quantity
20
Distribution
Distributed to the states of AZ, CA, CO, CT, FL, GA, ID, LA, MA, MO, MS, NC, OH, PA, TN, TX, & WA.
Lot, serial or model codes
Lot No. 05212015
510(k) numbers
["K112507"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3005985723
Firm city
Davie
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register