FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc.: Dimension CREA (Creatinine) and Dimension Vista CREA (Creatinine); The CREA method used on the Dimension clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of creatinine in human serum, plasma and urine.
- Recall number
- Z-2741-2015
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Dimension CREA (Creatinine) and Dimension Vista CREA (Creatinine); The CREA method used on the Dimension clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of creatinine in human serum, plasma and urine.
- Status
- Terminated
- Initiated
- 2015-07-15
- Posted
- 2015-09-05
- Terminated
- 2017-03-24
- Reason
- Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.
- Root cause
- Device Design
- Action
- Siemens sent an Urgent Medical Device Correction letter, dated July 2015, was provided to all Dimension CREA and Dimension Vista CREA customers who received the products to notify them of the issue, the potential risk to health, and actions to be taken by customers. If you have any questions please contact your Siemens Customer Care Center Technical Solutions at (800) 441-9250, or your local Siemens Technical Support Representative.
- Quantity
- Total number Dimension Vista - 32,165, Dimension - 238,634
- Distribution
- Worldwide Distribution - US (nationwide) and to the countries of : Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia Herzegovinia, Bulgaria, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Great Britain, Greece, Guatemala, Hungary, Iraq, Ireland, Honduras, Hong Kong, Iceland, Indonesia, India, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Libya, Macedonia, Malaysia, Mali, Mexico, Myanmar, Morocco, Netherlands, Nepal, Nicaragua, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkmenistan, United Arab Emirates, Uruguay, Venezuela, Vietnam and Yemen.
- Lot, serial or model codes
- Dimension CREA Lot Numbers: DB5195 FA5211 EA5233 DA5258 FA5275 BA5293 EA5315 EA5316 GA5342 FA5351 FA5352 FA6006 BA6026 BA6027 EB6047 GB6076 GA6097 BA6111 BA6112 FB6139 with expiration dates: 2015-07-14 2015-07-30 2015-08-21 2015-09-15 2015-10-02 2015-10-20 2015-11-11 2015-11-12 2015-12-08 2015-12-17 2015-12-18 2016-01-06 2016-01-26 2016-01-27 2016-02-16 2016-03-16 2016-04-06 2016-04-20 2016-04-21 2016-05-20 Dimension Vista CREA Lot Numbers: 14177BB 14212BD 14234BH 14261BC 14280BA 14301BB 14321AC 14346AC 15012AA 15042BA 15054AB 15076BA 15091BA 15125BB 15125BA 15141AA with expiration dates: 2015-06-26 2015-07-31 2015-08-22 2015-09-18 2015-10-07 2015-10-28 2015-11-17 2015-12-12 2016-01-12 2016-02-11 2016-02-23 2016-03-16 2016-03-31 2016-05-04 2016-05-04 2016-05-22
- 510(k) numbers
- ["K061238","K925668"]
- PMA numbers
- not on file
- Product code
- JFY
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28