Skip to the content

FDA Medical Device Recalls (openFDA)

bioMerieux, Inc.: NucliSens EasyQ Enterovirus serotype 68 v1.1: Catalog number: 200300 In Vitro Diagnostic Test

Recall number
Z-2740-2011
Recalling firm
bioMerieux, Inc.
Product
NucliSens EasyQ Enterovirus serotype 68 v1.1: Catalog number: 200300 In Vitro Diagnostic Test
Status
Terminated
Initiated
2011-05-10
Posted
2011-07-06
Terminated
2011-10-06
Reason
The NucliSENS EasyQ Enterovirus Reagents are unable to detect Enterovirus serotype 68, in respiratory samples.
Root cause
Device Design
Action
BioMerieux Vitek, Inc. sent an "Urgent Product Correction Notice" letter on/about May 10, 2011. The letter included; a description of the product, problem, possible impact to the patient and recommendations. Recommendations included instructing customers to discontinue the use of the kit for the detection of Enterovirus 68 and Rhinovirus 87 serotypes and noted the corrected IFU will be available the first week of April 2011. An Acknowledgement form was attached to be returned via fax. For additional questions please contact your local BioMerieux Clinical Customer representative at (800) 682-2666, option 3, option 2.
Quantity
42 kits
Distribution
(USA) Nationwide distribution including the states of MD, MI, MN, MT, NY and PA
Lot, serial or model codes
Catalog number: 200300, Lot number: 10042001, 9081001, 9112403, 9030902 and 9030901.
510(k) numbers
["K093383"]
PMA numbers
not on file
Product code
OOI
Firm FEI
3002769706
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register