Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Perfusion Systems: Medtronic DLP One-Piece Pediatric Arterial Cannula, Product Number/CFN 77010; Cardiopulmonary bypass vascular cannula

Recall number
Z-2738-2026
Recalling firm
Medtronic Perfusion Systems
Product
Medtronic DLP One-Piece Pediatric Arterial Cannula, Product Number/CFN 77010; Cardiopulmonary bypass vascular cannula
Status
Open, Classified
Initiated
2026-05-25
Posted
2026-07-15
Terminated
not on file
Reason
Medtronic received two complaints of loose foreign material confirmed as containing trace levels of blood.
Root cause
Packaging
Action
Medtronic issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 6/4/2026 via courier. The notice explained the problem, potential risk, and requested the product be discarded in a clearly marked biohazard bag or container and discarded. Consignees were directed to complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com and contact Medtronic Customer Service at 1-800-854-3570, Option 1 then Option 4, and reference this communication to initiate to arrange for credit or replacement product. Consignees were requested to share within their organization and If the product was forwarded to another facility, to notify those to whom the affected product was transferred.
Quantity
not on file
Distribution
Worldwide - US Nationwide distribution in the states of MD, CA, MI, WA, PA, NC, FL, AR.
Lot, serial or model codes
MATNR A7630009466801/UDI/DI 20763000946689, Lot Numbers: 0231650870, C232227220
510(k) numbers
["K024069","K180456"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000116158
Firm city
Brooklyn Park
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register