Skip to the content

FDA Medical Device Recalls (openFDA)

Agfa N.V.: DR 800. Digital Radiography X-ray System.

Recall number
Z-2738-2024
Recalling firm
Agfa N.V.
Product
DR 800. Digital Radiography X-ray System.
Status
Open, Classified
Initiated
2024-07-29
Posted
2024-08-27
Terminated
not on file
Reason
Potential for the front lever chain of the DR 800 table to fail.
Root cause
Device Design
Action
Consignees were sent an URGENT FIELD SAFETY NOTICE by email, dated 7/29/24. In the notice, consignees are asked to check the correct tightening of the floor plate fixing bolts if a high and persistent noise is heard from the device during the tilting or elevation movement of the tabletop. If the noise persists, the device must be stopped immediately in a horizontal position at the maximum height. Agfa will coordinate with consignees to schedule an inspection of affected devices and to install hardware intended to reinforce equipment before 10/15/24. The recall notification is to be shared within consignee organizations and forwarded to those devices were transferred to. Consignees are asked to maintain awareness of this notice until devices are corrected. Completed response forms are to be provided via email to cassandra.mcgowan@agfa.com. Consignees with any questions are to call 1-877-777-2432 and reference PRB2001106.
Quantity
141 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.
Lot, serial or model codes
Type No. 6010/200; UDI-DI: 05414904251805; Device Serial No. 925, 945, 948, 949, 990, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, 1404, 1453, 1460, 1489, 1506, 1532, 1568, 1704, 1733, 1757, 1888, 1893, 1950, 1954, 1957, 2010, 2070, 2130, 2244, 2297, 2371, 2378, 2901, 2931, 3068, 3217, 1507, 643, 763, 903, 915, 930, 969, 984, 992, 1032, 1042, 1049, 1052, 1123, 1148, 1197, 1217, 1323, 1341, 1203, 1279, 1458, 1509, 1521, 1627, 1676, 1691, 1756, 1828, 1850, 1878, 1921, 1924, 1926, 1932, 1938, 1941, 1943, 1956, 2079, 2088, 2094, 2101, 2115, 2129, 2131, 2157, 2169, 2192, 2193, 2225, 2233, 2243, 2251, 2253, 2254, 2277, 2289, 2310, 2314, 2318, 2333, 2340, 2359, 2377, 2409, 2415, 2416, 2428, 2436, 2437, 2449, 2450, 2475, 2483, 2513, 2538, 2580, 2642, 2678, 2687, 2756, 2781, 2833, 2842, 2873, 2886, 2892, 2941, 2955, 2992, 3101, 3134, 3143, 3151, 3161, 1920, 3006, 3040; Table Serial No. 32497, 34673, 34695, 34713, 34717, 34976, 34990, 35350, 36241, 36267, 36462, 37150, 37177, 38592, 38595, 38624, 38648, 39729, 39763, 39773, 39796, 41160, 41179, 41192, 42084, 42088, 42139, 42138, 43208, 43239, 43250, 44203, 45270, 45274, 46108, 46130, 49894, 51360, 52275, 53884, 39755, 31908, 32486, 32977, 34257, 34353, 34355, 34678, 34957, 34970, 34979, 35365, 35371, 38591, 35756, 36246, 36442, 36883, 37137, 37869, 37909, 36889, 37221, 38630, 39739, 39757, 40634, 41132, 41147, 41191, 42066, 42071, 42076, 42098, 42101, 42102, 42113, 42115, 42117, 42118, 42140, 43259, 43265, 43269, 43288, 44180, 44188, 44192, 44212, 44233, 44253, 44254, 45219, 45227, 45235, 45259, 45249, 45264, 45260, 45269, 45282, 45283, 45285, 45299, 45304, 46076, 46113, 46145, 46148, 46156, 46166, 46167, 47188, 46175, 47186, 47204, 47207, 47227, 47241, 47273, 48349, 48361, 48376, 48413, 49808, 49830, 49834, 49869, 49882, 49887, 51364, 51381, 51422, 52303, 52322, 52327, 52333, 52337, 42097, 51428, 52255.
510(k) numbers
K180589, K183275
PMA numbers
not on file
Product code
JAA
Firm FEI
3001556265
Firm city
Mortsel
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28