FDA Medical Device Recalls (openFDA)
Agfa N.V.: DR 800. Digital Radiography X-ray System.
- Recall number
- Z-2738-2024
- Recalling firm
- Agfa N.V.
- Product
- DR 800. Digital Radiography X-ray System.
- Status
- Open, Classified
- Initiated
- 2024-07-29
- Posted
- 2024-08-27
- Terminated
- not on file
- Reason
- Potential for the front lever chain of the DR 800 table to fail.
- Root cause
- Device Design
- Action
- Consignees were sent an URGENT FIELD SAFETY NOTICE by email, dated 7/29/24. In the notice, consignees are asked to check the correct tightening of the floor plate fixing bolts if a high and persistent noise is heard from the device during the tilting or elevation movement of the tabletop. If the noise persists, the device must be stopped immediately in a horizontal position at the maximum height. Agfa will coordinate with consignees to schedule an inspection of affected devices and to install hardware intended to reinforce equipment before 10/15/24. The recall notification is to be shared within consignee organizations and forwarded to those devices were transferred to. Consignees are asked to maintain awareness of this notice until devices are corrected. Completed response forms are to be provided via email to cassandra.mcgowan@agfa.com. Consignees with any questions are to call 1-877-777-2432 and reference PRB2001106.
- Quantity
- 141 units
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.
- Lot, serial or model codes
- Type No. 6010/200; UDI-DI: 05414904251805; Device Serial No. 925, 945, 948, 949, 990, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, 1404, 1453, 1460, 1489, 1506, 1532, 1568, 1704, 1733, 1757, 1888, 1893, 1950, 1954, 1957, 2010, 2070, 2130, 2244, 2297, 2371, 2378, 2901, 2931, 3068, 3217, 1507, 643, 763, 903, 915, 930, 969, 984, 992, 1032, 1042, 1049, 1052, 1123, 1148, 1197, 1217, 1323, 1341, 1203, 1279, 1458, 1509, 1521, 1627, 1676, 1691, 1756, 1828, 1850, 1878, 1921, 1924, 1926, 1932, 1938, 1941, 1943, 1956, 2079, 2088, 2094, 2101, 2115, 2129, 2131, 2157, 2169, 2192, 2193, 2225, 2233, 2243, 2251, 2253, 2254, 2277, 2289, 2310, 2314, 2318, 2333, 2340, 2359, 2377, 2409, 2415, 2416, 2428, 2436, 2437, 2449, 2450, 2475, 2483, 2513, 2538, 2580, 2642, 2678, 2687, 2756, 2781, 2833, 2842, 2873, 2886, 2892, 2941, 2955, 2992, 3101, 3134, 3143, 3151, 3161, 1920, 3006, 3040; Table Serial No. 32497, 34673, 34695, 34713, 34717, 34976, 34990, 35350, 36241, 36267, 36462, 37150, 37177, 38592, 38595, 38624, 38648, 39729, 39763, 39773, 39796, 41160, 41179, 41192, 42084, 42088, 42139, 42138, 43208, 43239, 43250, 44203, 45270, 45274, 46108, 46130, 49894, 51360, 52275, 53884, 39755, 31908, 32486, 32977, 34257, 34353, 34355, 34678, 34957, 34970, 34979, 35365, 35371, 38591, 35756, 36246, 36442, 36883, 37137, 37869, 37909, 36889, 37221, 38630, 39739, 39757, 40634, 41132, 41147, 41191, 42066, 42071, 42076, 42098, 42101, 42102, 42113, 42115, 42117, 42118, 42140, 43259, 43265, 43269, 43288, 44180, 44188, 44192, 44212, 44233, 44253, 44254, 45219, 45227, 45235, 45259, 45249, 45264, 45260, 45269, 45282, 45283, 45285, 45299, 45304, 46076, 46113, 46145, 46148, 46156, 46166, 46167, 47188, 46175, 47186, 47204, 47207, 47227, 47241, 47273, 48349, 48361, 48376, 48413, 49808, 49830, 49834, 49869, 49882, 49887, 51364, 51381, 51422, 52303, 52322, 52327, 52333, 52337, 42097, 51428, 52255.
- 510(k) numbers
- K180589, K183275
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 3001556265
- Firm city
- Mortsel
- Firm state
- not on file
- Firm country
- Belgium
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28