Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids.

Recall number
Z-2738-2014
Recalling firm
Baxter Healthcare Corp.
Product
SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids.
Status
Terminated
Initiated
2014-09-03
Posted
2014-10-01
Terminated
2017-03-16
Reason
One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014.
Root cause
Equipment maintenance
Action
Initial notification was initiated via telephone call by the Baxter Medina Service Center on 8/28/14 to all affected Sigma SPECTRUM Infusion Pump consignees. URGENT DEVICE CORRECTION Letters (dated 9/03/2014) were sent to the consignees via USPS first class mail on 9/03/14 formally informing them of the recall. The letters identified the affected product, the description of the issue, the hazard involved, as well as, the actions to be taken by customer. Customers are being instructed to immediately remove the pump from use and return the pump to Baxter for inspection. The firm will provide replacements. Customers are to contact Baxter Healthcare Medina at 800-356-3454 for technical questions regarding the letter.
Quantity
USA: 56 units, Canada: 2 units
Distribution
Worldwide Distribution -- USA and Canada.
Lot, serial or model codes
Software Versions 5.02.06, 6.02.06, and 6.02.11; Affected Serial Numbers: 712090, 723687, 723842, 724966, 725820, 735977, 751130, 752124, 755174, 768538, 771990, 774743, 781406, 783736, 784698, 794466, 805797, 808901, 809113, 812462, 815932, 842120, 857926, 858672, 862085, 863721, 865630, 873459, 889597, 890260, 903748, 912619, 920574, 920589, 956346, 957394, 965402, 966684, 976931, 977550, 978361, 983979, 984066, 984129, 984475, 985946, 987538, 993445, 995291, 996014, 996389, 1013037, 1004377, 1014565, 1014962, 967887, 950671, and 938428.
510(k) numbers
["K042121"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register