FDA Medical Device Recalls (openFDA)
Maquet Datascope Corp - Cardiac Assist Division: CS 100 Intra-Aortic Balloon Pump
- Recall number
- Z-2737-2017
- Recalling firm
- Maquet Datascope Corp - Cardiac Assist Division
- Product
- CS 100 Intra-Aortic Balloon Pump
- Status
- Terminated
- Initiated
- 2017-06-16
- Posted
- 2017-06-23
- Terminated
- 2019-11-29
- Reason
- The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
- Root cause
- Component design/selection
- Action
- Datascope Corp/MAQUET issued a press release on June 16, 2017. Datascope Corp/MAQUET sent an Urgent Product Recall Letter on June 19, 2017, to their customers via FedEx next day delivery service. Pursuant to the WARNINGS section of our CS100i, CS100 or CS300 IABP Operating/User Instructions, clinicians are instructed not to leave the patient unattended during IABP therapy. Until the service is performed, we recommend the IABP be powered on prior to initiation of therapy to allow the IABP to successfully complete its self-test. This action will take less than 60 seconds to perform. In the event the IABP fails to successfully complete the self-test and exhibits electrical test failure code 58, please remove the IABP from service and contact your local Maquet/Getinge Sales & Service Office. Customers will be instructed to complete and return the attached Medical Device Field Correction Response Form via fax to 1-973-807-9217 or send via email to IABP2017@getinge.com.ABP2017@getinge.com. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. For U.S. customers with technical questions, please contact our Technical Support Department (at 1-888-627-8383 and press 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST
- Quantity
- 12,319 units total
- Distribution
- Worldwide Distribution - (in over 100 countries) US (Nationwide)
- Lot, serial or model codes
- 0998-XX-3013-XX, 0998-UC-3013-XX
- 510(k) numbers
- ["K031636"]
- PMA numbers
- not on file
- Product code
- DSP
- Firm FEI
- 3001418283
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28