Philips Medical Systems (Cleveland) Inc: BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)
- Recall number
- Z-2736-2016
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)
- Status
- Terminated
- Initiated
- 2016-05-25
- Posted
- not on file
- Terminated
- 2018-07-02
- Reason
- The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
- Root cause
- Software design
- Action
- The firm is informing the consignee of this issue through Customer Information Letter.
- Quantity
- 424 units
- Distribution
- Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
- Lot, serial or model codes
- 6000016, 6000111, 6000114, 6000122, 6000002, 6000003, 6000004, 6000005, 6000006, 6000010, 6000011, 6000012, 6000013, 6000015, 6000017, 6000018, 6000019, 6000020, 6000022, 6000023, 6000024, 6000025, 6000026, 6000027, 6000028, 6000029, 6000030, 6000031, 6000032, 6000033, 6000035, 6000036, 6000037, 6000038, 6000040, 6000041, 6000042, 6000043, 6000044, 6000045, 6000047, 6000048, 6000049, 6000050, 6000051, 6000052, 6000053, 6000054, 6000055, 6000056, 6000057, 6000058, 6000059, 6000060, 6000061, 6000062, 6000063, 6000064, 6000065, 6000066, 6000067, 6000068, 6000069, 6000070, 6000071, 6000072, 6000073, 6000074, 6000075, 6000076, 6000077, 6000078, 6000079, 6000080, 6000081, 6000083, 6000084, 6000085, 6000086, 6000087, 6000088, 6000089, 6000090, 6000091, 6000092, 6000093, 6000094, 6000095, 6000096, 6000097, 6000098, 6000099, 6000100, 6000101, 6000102, 6000103, 6000104, 6000105, 6000106, 6000107, 6000108, 6000109, 6000110, 6000112, 6000113, 6000115, 6000116, 6000117, 6000118, 6000119, 6000120, 6000121, 6000123, 6000125, 6000126, 6000127, 6000128, 6000129, 6000130, 6000131, 6000132, 6000133, 6000134, 6000135, 6000136, 6000138, 6000139, 6000140, 6000141, 6000142, 6000143, 6000144, 6000145, 6000146, 6000147, 6000148, 6000149, 6000150, 6000151, 6000152, 6000153, 6000154, 6000155, 6000156, 6000157, 6000158, 6000159, 6000160, 6000161, 6000162, 6000164, 6000165, 6000166, 6000167, 6000168, 6000169, 6000170, 6000171, 6000172, 6000173, 6000174, 6000175, 6000176, 6000177, 6000178, 6000179, 6000180, 6000181, 6000182, 6000183, 6000184, 6000185, 6000186, 6000187, 6000188, 6000189, 6000190, 6000191, 6000192, 6000193, 6000194, 6000195, 6000196, 6000197, 6000198, 6000199, 6000200, 6000201, 6000202, 6000203, 6000204, 6000205, 6000206, 6000207, 6000208, 6000209, 6000210, 6000211, 6000212, 6000213, 6000214, 6000215, 6000216, 6000217, 6000218, 6000219, 6000220, 6000221, 6000222, 6000223, 6000224, 6000225, 6000226, 6000227, 6000228, 6000229, 6000230, 6000231, 6000232, 6000233, 6000234, 6000235, 6000236, 6000237, 6000238, 6000239, 6000240, 6000241, 6000242, 6000244, 6000245, 6000246, 6000247, 6000249, 6000250, 6000251, 6000252, 6000253, 6000254, 6000255, 6000256, 6000257, 6000258, 6000259, 6000260, 6000261, 6000262, 6000263, 6000264, 6000265, 6000266, 6000267, 6000268, 6000269, 6000270, 6000271, 6000272, 6000273, 6000274, 6000275, 6000276, 6000277, 6000278, 6000279, 6000280, 6000281, 6000283, 6000284, 6000285, 6000286, 6000287, 6000288, 6000289, 6000290, 6000291, 6000292, 6000293, 6000294, 6000295, 6000296, 6000297, 6000298, 6000299, 6000300, 6000301, 6000302, 6000303, 6000304, 6000305, 6000306, 6000307, 6000308, 6000309, 6000310, 6000311, 6000312, 6000313, 6000314, 6000315, 6000316, 6000317, 6000318, 6000319, 6000320, 6000321, 6000322, 6000323, 6000324, 6000325, 6000326, 6000327, 6000328, 6000329, 6000330, 6000331, 6000332, 6000333, 6000334, 6000335, 6000336, 6000337, 6000338, 6000339, 6000340, 6000341, 6000342, 6000343, 6000344, 6000345, 6000346, 6000347, 6000348, 6000349, 6000350, 6000351, 6000352, 6000353, 6000354, 6000355, 6000356, 6000357, 6000358, 6000359, 6000360, 6000361, 6000362, 6000363, 6000364, 6000365, 6000366, 6000367, 6000368, 6000369, 6000370, 6000371, 6000372, 6000373, 6000374, 6000375, 6000376, 6000377, 6000378, 6000379, 6000380, 6000381, 6000382, 6000383, 6000384, 6000385, 6000386, 6000387, 6000388, 6000389, 6000390, 6000391, 6000392, 6000393, 6000394, 6000395, 6000396, 6000398, 6000399, 6000400, 6000401, 6000402, 6000403, 6000404, 6000405, 6000406, 6000407, 6000408, 6000409, 6000410, 6000411, 6000412, 6000413, 6000414, 6000415, 6000416, 6000417, 6000418, 6000419, 6000420, 6000421, 6000422, 6000423, 6000424, 6000425, 6000426, 6000427, 6000428, 6340008, 6340009, 6340010, 6340011, 6000004A, 6000014, 6000039, 6000046, 6000082, 6000163, 6000243, 6000282, 6000006A
- 510(k) numbers
- ["K080927"]
- PMA numbers
- not on file
- Product code
- KPS
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28