Medtronic Perfusion Systems: MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3, TL7M46R7, TL2Q68R17, 7U65R2, M074704B, M230002A, BB7M38Rll, M450013B, BB8N86R4, HY2U43R9, 8C72R4, M650123A, HY8R55R ,TL7M46R7, M971406C, M450013B, 7U65R2, M394415A, M450013B. Enables the perfusionist to manage the patient's hematocrit and fluid status.

Recall number
Z-2736-2015
Recalling firm
Medtronic Perfusion Systems
Product
MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3, TL7M46R7, TL2Q68R17, 7U65R2, M074704B, M230002A, BB7M38Rll, M450013B, BB8N86R4, HY2U43R9, 8C72R4, M650123A, HY8R55R ,TL7M46R7, M971406C, M450013B, 7U65R2, M394415A, M450013B. Enables the perfusionist to manage the patient's hematocrit and fluid status.
Status
Terminated
Initiated
2015-08-18
Posted
2015-09-04
Terminated
2016-05-09
Reason
Hemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification. These potentially affected units have been distributed as stand-alone devices and in specific lots of Medtronic Perfusion Tubing packs.
Root cause
Nonconforming Material/Component
Action
Consignees were sent a Medtronic "Urgent Medical Device Recall" letter dated August 2015. The letter was addressed to "Risk Manager". The letter described the problem and the product involved in the recall. Informed consignees that no action is needed for product that has been used, but patients should continue to be managed in accordance with their standard patient management protocol. For unused product they recommend to immediately quarantine and return the affected product to Medtronic. Contact Customer Service at 800-854-3570 and reference RADAR #253277 to initiated a return and credit of product. Requested consignees to complete the Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality. If the product has been forwarded to another facility, consignees are advised to notify that facility. For questions contact Medtronic Lifeline Technical Services at 877-526-7890 or your Medtronic representative.
Quantity
Total 1594 (1306 US, 288 OUS)
Distribution
Worldwide Distribution-US (nationwide) including the states of AL, AR, CA, CO, FL, HI, ID, IN, KS, LA, MA, MS, MO, NY, NJ, OH, PA, TN, TX, UT, and WI, and the countries of Bolivia, Denmark, Dominican Republic, Italy, Latvia, Lebanon, Mexico, Panama, Singapore, Venezuela, and Zambia.
Lot, serial or model codes
209050007 209087876 209107465 209135076 209162502 209195930 209241496 209241613 209270018 209270092 209297994 209298091 209298094 209310403 209326025 209373545 209388464 209396017 209401269 209401312 209415313 209431890 209464644 209512240 209556873 209621268 209644548 209653827
510(k) numbers
["K012538","K122811","K800178","K891687"]
PMA numbers
not on file
Product code
DWE
Firm FEI
1000116158
Firm city
Brooklyn Park
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register