Beckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

Recall number
Z-2735-2024
Recalling firm
Beckman Coulter, Inc.
Product
DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Status
Open, Classified
Initiated
2024-05-22
Posted
2024-08-27
Terminated
not on file
Reason
When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.
Root cause
Software design
Action
The recalling firm issued letters dated 5/16/2024 on 5/22/2024 via postal mail and email notification. The letter described the issue, impact, action, and resolution. Beckman Coulter will implement the correction with a future software release and the consignee will be contacted by their Beckman Coulter service representative to schedule the software upgrade. If the device has been forwarded to another laboratory, the consignee is requested to provide a copy of the letter to that location. A response was requested within 10 days electronically via email or through completion and return of the enclosed Response Form.
Quantity
183 analyzers
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom of Great Britain and Northern Ireland.
Lot, serial or model codes
Software versions 1.17.0 or below, UDI-DI 15099590732103.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJE
Firm FEI
2122870
Firm city
Chaska
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register