Beckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
- Recall number
- Z-2735-2024
- Recalling firm
- Beckman Coulter, Inc.
- Product
- DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
- Status
- Open, Classified
- Initiated
- 2024-05-22
- Posted
- 2024-08-27
- Terminated
- not on file
- Reason
- When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.
- Root cause
- Software design
- Action
- The recalling firm issued letters dated 5/16/2024 on 5/22/2024 via postal mail and email notification. The letter described the issue, impact, action, and resolution. Beckman Coulter will implement the correction with a future software release and the consignee will be contacted by their Beckman Coulter service representative to schedule the software upgrade. If the device has been forwarded to another laboratory, the consignee is requested to provide a copy of the letter to that location. A response was requested within 10 days electronically via email or through completion and return of the enclosed Response Form.
- Quantity
- 183 analyzers
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom of Great Britain and Northern Ireland.
- Lot, serial or model codes
- Software versions 1.17.0 or below, UDI-DI 15099590732103.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2122870
- Firm city
- Chaska
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28