FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Medfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1

Recall number
Z-2734-2020
Recalling firm
Smiths Medical ASD Inc.
Product
Medfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1
Status
Open, Classified
Initiated
2020-06-26
Posted
2020-08-06
Terminated
not on file
Reason
Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
Root cause
Software design
Action
Consignees are being contacted via direct mailing and email. Consignees are asked to review and complete the Recall Response form indicating understanding of notification. Distributors are asked to forward the notification to parties to whom product has been distributed. Smiths Medical will send out 2 additional notifications by email. Phone calls will be made to non-responders.
Quantity
35616 units
Distribution
Worldwide distribution - United Arab Emirates, United States, Afghanistan, Albania, Australia, Belgium, Bermuda, Canada, China, Germany, Denmark, Great Britain, Hong Kong, Indonesia, Ireland, India, Korea, Kuwait, Lebanon, Mexico, Malaysia, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Saudi Arabia, Trinidad and Tobago
Lot, serial or model codes
Model Numbers: 4000-0105-249, 4000-0105-50, 4000-0105-51, 4000-0105-78, 4000-0106-00, 4000-0106-01, and 4000-0106-231. Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1
510(k) numbers
["K040899","K111386"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register