FDA Medical Device Recalls (openFDA)
Encore Medical, Lp: Foundation Knee System Instrumentation, Tibial Broach Assembly. Used in conjunction with the Tibial Sizing Template and Tibial Broach Guide to prepare the resected tibial plateau for acceptance of a keeled tibial baseplate stem.
- Recall number
- Z-2734-2011
- Recalling firm
- Encore Medical, Lp
- Product
- Foundation Knee System Instrumentation, Tibial Broach Assembly. Used in conjunction with the Tibial Sizing Template and Tibial Broach Guide to prepare the resected tibial plateau for acceptance of a keeled tibial baseplate stem.
- Status
- Terminated
- Initiated
- 2011-04-21
- Posted
- 2011-07-05
- Terminated
- 2012-04-06
- Reason
- Strike plate may dislodge from the broach stem handle during impaction of the device.
- Root cause
- Process control
- Action
- Encore Medical, Lp (d.b.a. DJO Global) sent out a "Urgent Field Safety Notice" dated April 20, 2011 to all affected customers. The firm also e-mailed the consignees as well. The letter described the product, affected lots, reason for recall and provided recommendations including; asking customers contact firm for a replacement, provided a "Field Safety Response Form" to fill out and return. The firm also e-mailed the consignees as well. For additional information please contact the Director of Commercial Logistics at (512) 834-6330 or the International Customer Service at 512 834-6275.
- Quantity
- 15 units
- Distribution
- Louisiana and Arkansas and the countries of Germany, Italy, United Kingdom and Japan.
- Lot, serial or model codes
- P/N 801-01-013 Lot 37641L01
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HTQ
- Firm FEI
- 1000116912
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28