Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

Recall number
Z-2733-2015
Recalling firm
Beckman Coulter Inc.
Product
Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.
Status
Terminated
Initiated
2015-07-28
Posted
2015-09-03
Terminated
2016-04-13
Reason
Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.
Root cause
Under Investigation by firm
Action
On 07/28/2015 the firm sent out customer notification letters via first class mail and e-mail. The firm's letter contained the following instructions: 1. Discard all Access Thyroglobulin reagent pack lots listed in the letter. 2. Evaluate Access Thyroglobulin results <10 ng/mL when using the reagent pack lot numbers listed to determine if controls and/or patients were higher than expected or did not match the clinical status of the patient. 3. At the discretion of the Laboratory Director, notify clinicians that it is possible the Access Thyroglobulin sample results reported by the laboratory were affected by this issue. The firm's notification letter requests that the information be shared with laboratory staff. If any affected products have been forwarded provide a copy of the letter to them. The firm requests that customers complete and return the Response Form within 10 days. Any questions contact the Customer Care Center: 1. From website at http://www.beckmancoulter.com 2. By phone: call 1-800-854-3633 in the United States and Canada. 3. Outside the United States and Canada, contact local Beckman Coulter representative
Quantity
3,035 (U.S.) and 7,123 (International)
Distribution
Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Republic of, Lebanon, Lithuania, Mexico, Morocco, New Zealand, Oman, Philippines, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam and Yemen.
Lot, serial or model codes
Catalogue number: 33860 Lot numbers: 431990 434727 430969 522896 523251  Expiration date: 10/31/16 12/31/16 08/31/16 04/30/17 05/31/17
510(k) numbers
["K002905","K031269"]
PMA numbers
not on file
Product code
MSW
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register