Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc. Endoscopy Division: DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew DYONICS Disposable Endoscopic Blades are indicated for resection of soft and osseous tissues in both large and small articular cavities.

Recall number
Z-2732-2011
Recalling firm
Smith & Nephew, Inc. Endoscopy Division
Product
DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew DYONICS Disposable Endoscopic Blades are indicated for resection of soft and osseous tissues in both large and small articular cavities.
Status
Terminated
Initiated
2011-05-06
Posted
2011-07-01
Terminated
2012-05-09
Reason
Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.
Root cause
Process control
Action
Smith & Nephew notified customers by an "URGENT - PRODUCT RECALL 1st NOTIFICATION" letter dated May 6, 2011. The letter describes the product, problem, and actions to be taken. The letter asked customers to call 800-343-5717 (option 3) or send an email to endo.andreturn@smith-nephew.com for a Return Authorization (RA) Number to return product, obtain credit or order replacement blades. Customers are to complete the form and return to the address provided being sure to reference the RA number. For any questions regarding this recall customers were instructed to call 508-337-3731.
Quantity
555 units
Distribution
Worldwide Distribution -- USA (nationwide) and countries of: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong Kong, Italy, Japan, Mexico, Netherlands, Turkey, and United Kingdom.
Lot, serial or model codes
Lot Number: 50561144
510(k) numbers
not on file
PMA numbers
not on file
Product code
HAB
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register