Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Corporation: Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-013-030DC

Recall number
Z-2727-2024
Recalling firm
Stryker Corporation
Product
Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-013-030DC
Status
Completed
Initiated
2024-07-24
Posted
2024-08-26
Terminated
not on file
Reason
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Root cause
Under Investigation by firm
Action
Stryker issued Urgent Medical Device Recall letter on 7/24/424 to Materials Manager, Risk Manager, OR Director via Fed'x. Letter states reason for recall, health risk, and action to take: 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A credit will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. If you have any questions about this matter, contact Stryker Instruments by phone (269) 568-1481 or email at Instruments.Recalls@stryker.com.
Quantity
2670 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.
Lot, serial or model codes
GTIN: 07613327501209 Lot Numbers: 22230017 22333017 22342017 23011017 23093017 23172017 23188017 23221017 23311017 24004027
510(k) numbers
["K210377"]
PMA numbers
not on file
Product code
HBE
Firm FEI
3015967359
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register