Skip to the content

FDA Medical Device Recalls (openFDA)

TELEFLEX MEDICAL INC: Rusch Greenlite Product Code: 0004550003 GTIN: 14026704663078 (each) - Product Usage: The laryngoscope is a device intended to allow direct visualization of the laryngeal structure and is used during tracheal intubation.

Recall number
Z-2727-2020
Recalling firm
TELEFLEX MEDICAL INC
Product
Rusch Greenlite Product Code: 0004550003 GTIN: 14026704663078 (each) - Product Usage: The laryngoscope is a device intended to allow direct visualization of the laryngeal structure and is used during tracheal intubation.
Status
Terminated
Initiated
2020-07-06
Posted
not on file
Terminated
2023-08-31
Reason
Breaking at the point of a welded joint
Root cause
Process control
Action
Teleflex issued an Urgent Medical Device Recall on 07/06/2020. It states the following: 1. Immediately discontinue distribution and quarantine any products with the product code and lot number listed. 2. Using the provided customer letter and recall acknowledgement form templates, communicate this recall to any of your customers who have received product included within the scope of the recall. 3. Have the customers return any affected product to you, together with a completed Recall Acknowledgement Form, for consolidation and return to Teleflex Medical. In the event that an alternative approach is needed, contact Teleflex Medical Customer Service for more information at 1-866-396-2111. 4. To return recalled products to Teleflex Medical, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow the firm to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical. 5. Once you have completed returning all of the recalled products from your own inventory, and collecting and consolidating all of the recalled products from your customers, please check the box on the enclosed Recall Acknowledgment Form that indicates that you have completed the recall and fax it to 1-855-419-8507, Attn: Customer Service, or email it to recalls@teleflex.com. This will allow to document completion of the recall. 6. If you and your customers have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow to document your receipt of this letter. If you have any other questions,contact your local sales representative or Customer Service at 1-866-396-2111.
Quantity
27341 units
Distribution
US Nationwide distribution including in the states of FL, KY, CA, ND, UT, MA, MD, AZ, MS, TX, MI, WV, AL, MO, SC, VA, AR, PA, OK, NV, NM, NE, IN, IA, NY, GA, NJ, MN, IL, NC, NH, TN, ID, CO, HI, MT, OR, ME, CT, LA, OH, RI, SD, WA, DC, KS, AK.
Lot, serial or model codes
Lot/Batch Numbers: 1812431 1901431 1902431 1903431 1904431 1905431 1906431 1907431 1908431 1909431
510(k) numbers
not on file
PMA numbers
not on file
Product code
CCW
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register