Brentwood Medical Technology Corp: IQecg IQiC Interface, Part Numbers: 4-100-1407, Software Version 6.0 IQecg is to be used, under the supervision of a Physician, to obtain Electrocardiograms from the human body surface.

Recall number
Z-2724-2011
Recalling firm
Brentwood Medical Technology Corp
Product
IQecg IQiC Interface, Part Numbers: 4-100-1407, Software Version 6.0 IQecg is to be used, under the supervision of a Physician, to obtain Electrocardiograms from the human body surface.
Status
Terminated
Initiated
2010-05-18
Posted
2011-07-14
Terminated
2011-08-22
Reason
The recall was initiated because when a user of the IQmanager, IQdevices SDK, and IQiC (Centricity) software versions listed above accesses the QT dialogue box to either accept or edit the QT value, the QTc value is recalculated using a Hodges QTc equation, regardless of patient's age. The software does this without notifying the user of the change from use of the Bazett equation, which is utiliz
Root cause
Software design
Action
The firm, Midmark, sent an "URGENT: MEDICAL DEVICE FIELD CORRECTIVE ACTION" letter on May 19, 2010 with an attached Corrective Return Material Authorization (RMA) form via US Postal Service Certified Mail to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine their IQecg, IQmanager, or lQiC Centricity software to determine the software version currently in use. If it is a listed product, it is subject to this field corrective action. A sample software label was attached to aid the consignees/customers in locating and identifying the software CD. For customers who have already installed the software, the version number can be identified in the "About" page. Consignees/customers were instructed to complete the enclosed Field RMA form and return it with the affected software to: Quality Assurance Department Midmark Diagnostics Group 1125 West 190 S t. Gardena, CA 90248 In addition, if they further distributed this product, identify customers and notify them at once of this product field corrective action. Replacement software was to be shipped to the consignees/customers upon receipt of the returned software CD. If they require a UPS return shipping label or an on-line download of replacement software, contact Midmark Diagnostics Group Support Services at 800-624-8950 extension 5145 or 5108. If you have any questions, call Midmark Diagnostics Group Support Services at 800-624-8950 extension 5145 or QA/RA Manager, Midmark Diagnostics Group 310-516-5116.
Quantity
90 units
Distribution
Worldwide distribution: USA (Nationwide) and countries of: Algeria, Australia, Canada, France and Saudi Arabia.
Lot, serial or model codes
Software Version 6.0
510(k) numbers
["K955023"]
PMA numbers
not on file
Product code
DSI
Firm FEI
2081230
Firm city
Gardena
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register