Skip to the content

FDA Medical Device Recalls (openFDA)

Transonic Systems Inc: Transonic Flow Probe, 8 mm, Catalog Number HQC8FMV -- Product Usage: to measure flow intra-operatively.

Recall number
Z-2723-2015
Recalling firm
Transonic Systems Inc
Product
Transonic Flow Probe, 8 mm, Catalog Number HQC8FMV -- Product Usage: to measure flow intra-operatively.
Status
Terminated
Initiated
2015-07-27
Posted
2015-09-03
Terminated
2016-07-18
Reason
Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
Root cause
Software Manufacturing/Software Deployment
Action
Transonic sent an Urgent Medical Device Recall letters and Response Forms dated July 27, 2015 to customers via Certified Mail, email and/or fax. The letter identified the affected product, problem and actions to be taken. Customers are instructed to contact Transonic Systems' Customer Service at 1-800-353-3569 or 1-607-257-5300 to arrange for replacement (repair/exchange) immediately.
Quantity
Domestic: 5 units
Distribution
Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
Lot, serial or model codes
Serial Numbers: HQC8FMV2142, HQC8FMV2143, HQC8FMV2144, HQC8FMV2145, HQC8FMV2146
510(k) numbers
["K872048"]
PMA numbers
not on file
Product code
DPW
Firm FEI
1319030
Firm city
Ithaca
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register