Skip to the content

FDA Medical Device Recalls (openFDA)

Treace Medical Concepts, Inc.: Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Recall number
Z-2718-2024
Recalling firm
Treace Medical Concepts, Inc.
Product
Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.
Status
Open, Classified
Initiated
2024-06-02
Posted
2024-08-23
Terminated
not on file
Reason
Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.
Root cause
Under Investigation by firm
Action
Distributors were contacted via phone or in person to hold product starting on 06/02/2024. Recall notifications were disseminated on 06/13/2024 via email containing the recall letter, acknowledgement, and shipping label for return. On 07/29/2024, Treace Medical contacted surgeons via letter. Surgeons were instructed to monitor patients for adverse events and to complete and return the acknowledgement form.
Quantity
40 kits containing 40 plates
Distribution
US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.
Lot, serial or model codes
UDI: (01)00810111222464(17)290424(10)300542404
510(k) numbers
["K240173"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3011623994
Firm city
Ponte Vedra
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register