Skip to the content

FDA Medical Device Recalls (openFDA)

ARJOHUNTLEIGH POLSKA Sp. z.o.o.: ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.

Recall number
Z-2718-2016
Recalling firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Product
ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.
Status
Terminated
Initiated
2016-06-29
Posted
not on file
Terminated
2019-03-08
Reason
ArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees-28 degrees. Although such scenario has been defined as highly unlikely to occur during use with patients, as a commitment to quality, ArjoHuntleigh is providing customers with information on action needed. The Sara Combilizer may stay in use until the upgrade will be performed, on the condition that the tilt angle in the device ''stretcher mode" will be no greater than minus 15 degrees and the device is not loaded over 330 lbs (150 kg).
Root cause
Device Design
Action
A Field Safety Notice, a Customer Response Form and the Sara Combilizer Addendum to the Instructions For Use (IFU) will be mailed to all affected customers.
Quantity
106
Distribution
CO, AL, MA, NC, MI, NY, FL, MD, CA, VA, CT, TX, and IL; Australia, Austria, Denmark, Germany, Israel, Japan, Norway, Sweden, Switzerland, UAE, and UK.
Lot, serial or model codes
Device Listing Number E101987, Model Number APA1000-XX; Serial numbers  P0359671 , P0360103 , P0338345 , P0354548 , P0357690 , P0359672 , P0359673 , P0348548 , P0349629 , P0361336 , P0361337 , P0361875 , P0361876 , P0361877 , P0360104 , P0360105 , P0346000 , P0373592 , P0357691 , P0349634 , P0354549 , P0349627 , P0339257 , P0321458 , P0299662 , P0325930 , P0325931 , P0325933 , P0310854 , P0344175 , P0356357 , P0356362 , P0356363 , P0360990 , P0375493 , P0298175 , P0309635 , P0310848 , P0323622 , P0323624 , P0325167 , 1404000118 , P0296131 , P0296315 , P0298465 , P0319583 , P0333285 , P0333288 , P0339264 , P0343583 , P0343584 , P0343585 , P0343684 , P0344174 , P0353246 , P0364628 , 1404000112 , P0333284 , P0333290 , P0294982 , P0294983 , P0295762 , P0298158 , P0309472 , P0317162 , P0325970 , P0331310 , 1404000113 , 1404000119 , 1404000120 , P0325967 , P0332385 , P0333283 , P0339574 , P0296130 , P0296318 , P0321460 , P0332383 , 1404000115 , P0331312 , P0331311 , P0331318 , P0331319 , P0325968 , P0295846 , P0294981 , P0294985 , P0295847 , P0332386 , P0294984 , P0332384 , 1404000116 , 1404000117 , P0321461 , 1404000114 , P0294986 , P0298176 , P0323011 , P0361422 , P0361423 , P0363456 , P0357692 , P0360102 , P0328718 , P0328720 , P0328724 .
510(k) numbers
not on file
PMA numbers
not on file
Product code
INQ
Firm FEI
3007420694
Firm city
KOMORNIKI
Firm state
not on file
Firm country
Poland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register