Interventional Spine Inc: PERPOS PLS SYSTEM, CATALOG NO. 9045-01 A sterile, medical device kit consisting of the components listed in Table 1-1 below. The kit is used by medical professionals to stabilize the spine as an aid to fusion through immobilization of the facet joints. A Surgical Technique brochure is provided in Appendix A and other labeling, including instructions for use, is provided in Appendix B.
- Recall number
- Z-2718-2011
- Recalling firm
- Interventional Spine Inc
- Product
- PERPOS PLS SYSTEM, CATALOG NO. 9045-01 A sterile, medical device kit consisting of the components listed in Table 1-1 below. The kit is used by medical professionals to stabilize the spine as an aid to fusion through immobilization of the facet joints. A Surgical Technique brochure is provided in Appendix A and other labeling, including instructions for use, is provided in Appendix B.
- Status
- Terminated
- Initiated
- 2011-03-23
- Posted
- 2011-07-01
- Terminated
- 2013-02-14
- Reason
- This recall has been initiated due to K-wires with a distal tip diameter that may be greater than specified causing an interference fit with the canulated screw. Removing the K-wire during a procedure may dislocate the implant resulting in the implant loosening or being pulled out or position. Use of this product may result in a delay of the surgical procedure and/or pull-out of the facet screw
- Root cause
- Pending
- Action
- Interventional Spine sent an RECALL NOTIFICATION letter dated March 23, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. The customer was instructed to take the following actions: Immediately examine inventory and quarantine product subject to recall. In addition, if this product was further distributed, please notify the Vice President Quality Assurance at once. Consignee were asked to also complete and return the enclosed response form as soon as possible. Consignee with questions were directed to the Consignees' Regional Director or the Vice President Quality Assurance. For further questions, please call (949) 472-0006
- Quantity
- 864 units total for all products
- Distribution
- Nationwide Distribution including the states of to NC, NY, TN & VA
- Lot, serial or model codes
- LOT #'s: 040309-A, 041409-A
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HXI
- Firm FEI
- 3003523239
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28