Skip to the content

FDA Medical Device Recalls (openFDA)

KCI USA, Inc.: CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50

Recall number
Z-2717-2014
Recalling firm
KCI USA, Inc.
Product
CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50
Status
Terminated
Initiated
2014-05-19
Posted
2014-09-29
Terminated
2015-09-18
Reason
Blade can shift during shipping which makes the product ineffective.
Root cause
Device Design
Action
KCI sent an Urgent - Voluntary Medical Device Field Notification letter dated June 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is sending Field Notifications to affected customers instructing them to complete appropriate actions. Option I: Contact your local KCI Representative or KCI Customer Technical Support at 1-800-275-4524 and select Option 3 to arrange for return and replacement of the affected harvester(s). Option 2: If you prefer, you may re-set the blade after removal from the packaging by gently tapping the curved end of the harvester against the palm of your hand, as shown in Figure 3, until the blade is no longer visible in the harvester top plate, as shown in Figure 4.
Quantity
1,331 units
Distribution
US nationwide Distribution
Lot, serial or model codes
Part number CT-H25 with the following lot numbers: 3244680000, 3446210000, 3246280000, 3246480000, 3246490000, 3246970000, 3247020000, 3247030000, 3247130000, 3247440000, 3247670000, 3247680000, 3247980000, 3248070000, 3248200000, 3248300000, 3249120000, 3249130000, 3249140000, 3249150000, 3249160000.  Part number CT-H50 with the following lot numbers: 324782, 324796, 324797, 3245120000, 3245670000, 3245680000, 3246270000, 3246490000, 3246690000, 3246700000, 3246940000, 3246980000, 3247140000, 3247160000, 3247170000, 3247250000, 3247310000, 3247470000, 324780000, 3247810000, 3247820000, 3247960000, 3247970000, 3248080000, 3248190000, 3248230000, 3248300000, 3248380000, 3248960000, 3248970000, 3248980000, 3248990000, 3249000000, 3249010000, 3249020000, 3249030000, 3249040000, 3249050000, 3249060000, 3249070000, 3249080000, 3249090000, 3249100000, 3249110000, 3249390000, 3249580000.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GFD
Firm FEI
3005178245
Firm city
San Antonio
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28