Arjo, Inc. dba ArjoHuntleigh: ArjoHuntleigh Flowtron ACS900, medical pump, software V1.099 Device is used to prevent the occurrence of Deep Vein Thrombosis (DVT) in at-risk patients. The system must be used as part of a prescribed plan of car under the supervision of trained medical and/or Clinical staff. The system is intended for use only in professional healthcare facilities. It is not intended to be used in the home healthcare environment.

Recall number
Z-2714-2017
Recalling firm
Arjo, Inc. dba ArjoHuntleigh
Product
ArjoHuntleigh Flowtron ACS900, medical pump, software V1.099 Device is used to prevent the occurrence of Deep Vein Thrombosis (DVT) in at-risk patients. The system must be used as part of a prescribed plan of car under the supervision of trained medical and/or Clinical staff. The system is intended for use only in professional healthcare facilities. It is not intended to be used in the home healthcare environment.
Status
Terminated
Initiated
2017-06-14
Posted
2017-06-26
Terminated
2022-08-12
Reason
Reports of the device failure involving delivering continuous pressure to the calf garment and not deflating as intended, without triggering an alarm.
Root cause
Under Investigation by firm
Action
ArjoHuntleigh sent a Field Safety Notice and Customer Response form to affected customers describing the reason for the recall and the actions to be taken. The instruction to customers is to review the Field Safety Notice and complete the Customer Response Form to acknowledge receipt and understanding of the Field Safety Notice and the required actions. These actions taken are as follows: 1. Ensure that all caregivers and users of the ArjoHuntleigh Flowtron ACS900 pump are made aware of this Field Safety Notice (FSN) and all listed devices at facility are available to be upgraded to software version V2.000 during the service technician visit. 2. Complete and sign the enclosed Customer Response Form and return this form to the local ArjoHuntleigh office. Note: ArjoHuntleigh will contact the designated facility representative listed on the Customer Response Form to schedule service to upgrade the software on all listed devices at your facility, free of charge. The device may stay in use until the upgrade will be performed, on the condition that the patients limbs are monitored frequently and the garments are checked to ensure that they are correctly fitted to the patient and that deflation of the garments is occurring on a regular basis. It is also recommended that the pump LCD display is regularly checked to ensure that it shows that the garments are inflating and deflating correctly (in the limited number of complaints, the LCD display was blank). This Field Safety Notice must be distributed to those individuals who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Additionally they should maintain awareness of the notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. The letter states that if the customer has additional questions or they require assistance completing the Customer Response Form, they should contact ArjoHu
Quantity
40,675 units worldwide
Distribution
Worldwide Distribution - US (Nationwide) and internationally to the following countries: United Kingdom, Australia, New Zealand, Japan, Italy, United Arab Emirates, France, Canada, Brazil, Austria, Saudi Arabia, Qatar, India, Switzerland, Bahrain, Kuwait, South Africa, Lebanon, Netherlands, Poland, Spain, Ecuador, Hong-Kong, Colombia, Malaysia, Paraguay, Mauritius, Cyprus, Greece, Oman, Norway, Ireland, Pakistan, Czech Republic, Mexico, Sweden, Tanzania, Hungary, Singapore, Jordan, Israel, Denmark, Chile, and Germany.
Lot, serial or model codes
Model numbers: 526000-01, 526000-01P, 526000-02, 526000-02P. Serial numbers range from 1400028031 to 1600048470.
510(k) numbers
["K143438"]
PMA numbers
not on file
Product code
JOW
Firm FEI
1419652
Firm city
Addison
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register