Skip to the content

FDA Medical Device Recalls (openFDA)

SCC Soft Computer: SoftReports versions 1.1.6.x and 1.1.7.x SoftReports version 1.1.6 released 2/2008; Version 1.1.7 released 6/2008. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. SoftReports is a report designer and web-based report launching tool to be used by knowledgeable, trained, and experienced personnel. It provides the ability to create ad hoc, user defined queries, and reports using data from other SCC products. SoftReports enables users to customize standard reports generated by SCC systems to meet their specific needs. This system performs the following functions: 1. The ability to customize report layouts using beginner, intermediate, or advanced options. 2. The ability to import and export reports. 3. The ability to provide custom desktops to launch SQL's within other SCC applications. 4. The ability to preview and modify layouts specific to users' needs. 5. The ability to modify elements of reports using the comprehensive control formatting toolbar. 6. The ability to incorporate charts and graphics into reports. 7. The ability to modify existing query templates. 8. The ability to create independent and embedded queries. 9. The ability to pars SQL statements and notify users of errors within the syntax. 10. The ability to easily update criteria parameters to indicate input needed for executing queries. 11. The ability to view and print any report that is created using the Designer. 12. The ability to schedule, execute, and track ad-hoc reports. This system does not perform the following functions: 1. SoftReports is not configured to write back to the database. 2. SoftReports does not support user-created Ad Hoc reports being used for diagnostic purposes. 3. SoftReports is not intended to alter the meaning of patient result reports (test reports).

Recall number
Z-2714-2011
Recalling firm
SCC Soft Computer
Product
SoftReports versions 1.1.6.x and 1.1.7.x SoftReports version 1.1.6 released 2/2008; Version 1.1.7 released 6/2008. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. SoftReports is a report designer and web-based report launching tool to be used by knowledgeable, trained, and experienced personnel. It provides the ability to create ad hoc, user defined queries, and reports using data from other SCC products. SoftReports enables users to customize standard reports generated by SCC systems to meet their specific needs. This system performs the following functions: 1. The ability to customize report layouts using beginner, intermediate, or advanced options. 2. The ability to import and export reports. 3. The ability to provide custom desktops to launch SQL's within other SCC applications. 4. The ability to preview and modify layouts specific to users' needs. 5. The ability to modify elements of reports using the comprehensive control formatting toolbar. 6. The ability to incorporate charts and graphics into reports. 7. The ability to modify existing query templates. 8. The ability to create independent and embedded queries. 9. The ability to pars SQL statements and notify users of errors within the syntax. 10. The ability to easily update criteria parameters to indicate input needed for executing queries. 11. The ability to view and print any report that is created using the Designer. 12. The ability to schedule, execute, and track ad-hoc reports. This system does not perform the following functions: 1. SoftReports is not configured to write back to the database. 2. SoftReports does not support user-created Ad Hoc reports being used for diagnostic purposes. 3. SoftReports is not intended to alter the meaning of patient result reports (test reports).
Status
Terminated
Initiated
2010-06-23
Posted
2011-06-30
Terminated
2012-07-26
Reason
SCC Soft Computer is issuing a correction for SoftReports Versions 1.1.6.x released 02/2008 and 1.1.7.x released 06/2008. Potential hazard - when using SoftReports Layout Designer, results displayed in the report may not match the patient information in the report header.
Root cause
Software design
Action
SCC Soft Computers sent a "RISK TO HEALTH NOTIFICATION TASK" letter dated June 9, 2010 to affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were offered an upgrade or hot fix. The letter requested acknowledgement of receipt of the task notification and permission to install the hot fix.
Quantity
49
Distribution
Worldwide Distribution--USA (nationwide) and the country of Canada.
Lot, serial or model codes
Versions 1.1.6.x and 1.1.7.x.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LNX
Firm FEI
1000307651
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register