Nidek Inc: SPECULAR MICROSCOPE CEM-530; Software version 1.08 and 1.09. Opthalmic: The NIDEK Specular Microscope CEM-530 provides non-contact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images. The captured images and analysis results of endothelium are used for intraocular or corneal surgery, postoperative follow-up and corneal observation such as for endothelial disorders or the corneal state of patients who wear extended-wear contact lenses.

Recall number
Z-2711-2016
Recalling firm
Nidek Inc
Product
SPECULAR MICROSCOPE CEM-530; Software version 1.08 and 1.09. Opthalmic: The NIDEK Specular Microscope CEM-530 provides non-contact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images. The captured images and analysis results of endothelium are used for intraocular or corneal surgery, postoperative follow-up and corneal observation such as for endothelial disorders or the corneal state of patients who wear extended-wear contact lenses.
Status
Terminated
Initiated
2016-08-08
Posted
2016-09-01
Terminated
2017-01-09
Reason
Software version 1.08 and 1.09 for the Specular Microscope CEM 530 included a change of analysis results feature that was not reviewed and approved by the FDA.
Root cause
Software change control
Action
The firm, Nidek, sent a Dear Doctor letters entitled "RECALL NOTIFICATION" dated August 15, 2016 to affected sites by Federal Express. The letter informs users of the issue and stated that Nidek field service representatives will visit to install updated software. If you have any questions, please call Customer Service at 1-800-722-0219; Monday-Friday 8am to 5pm PST or email: cem530update@noritsuservice.com.
Quantity
29 affected devices
Distribution
US Distribution to states of: TX, KS, FL, NV, CA, PA, KY, OR, TN, AL, VA, MO, OK, CT, NY. LA, NM, and IN.
Lot, serial or model codes
Microscopes with serial numbers: 120004,120005, 120006, 120007, 120009, 120010, 120011,120012, 120013, 120014, 0120016, 120019, 120021, 120023, 120024, 120026, 120030, 120036, 0120038, 120039, 120040, 120042, 120044,120045, 120046, 120047, 120048, 120049, 120050.
510(k) numbers
["K151706"]
PMA numbers
not on file
Product code
NQE
Firm FEI
2936921
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register