FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: APLIO 300 TUS-A300 diagnostic ultrasound system

Recall number
Z-2710-2017
Recalling firm
Toshiba American Medical Systems Inc
Product
APLIO 300 TUS-A300 diagnostic ultrasound system
Status
Terminated
Initiated
2017-05-01
Posted
not on file
Terminated
2023-10-11
Reason
When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.
Root cause
Software Manufacturing/Software Deployment
Action
Toshiba Medical sent an Urgent Medical Device Correction letter on May 1, 2017. The consignee is instructed not to use the Continuous Trace method for cardiac velocity trace measurements to quantify the MPG values. Modified software is being provided. A Toshiba service representative will contact the consignee to make an appointment to install the modified software to fix the problem. Customers were asked to complete the form and fax it to the toll-free number shown at the top of the form. The form can also be sent by email to RAffairs@tams.com. Customers with questions were instructed to call 800-521-1968.
Quantity
90 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Serial Numbers: AAC1692086, AAC1692087, AAC1692088, AAC1692089, AAC1692090, AAC16Y2091, AAC16Y2092, AAC16Y2093, AAC16Y2094, AAC16Y2095, AAC16Y2096, AAC16Y2097, AAC16Y2098, AAC16Y2099, AAC16Y2100, AAD16Z2101, AAD16Z2102, AAD16Z2103, AAD16Z2104, AAD16Z2105, AAD1712106, AAD1712107, AAD1712108, AAD1712109, AAD1712110, AAD1712111, AAD1712112, AAD1722114, AAD1722115, AAD1722116, AAD1722117, AAA1652011, AAA1652012, AAA1652013, AAA1652014, AAA1652015, AAA1672026, AAA1672027, AAA1672028, AAA1672029, AAA1672030, AAA1672031, AAA1672032, AAA1672033, AAA1672034, AAA1672035, AAA1672036, AAA1672037, AAA1672038, AAA1672040, AAA1682041, AAA1682042, AAA1682043, AAA1682044, AAA1682045, AAA1682046, AAA1682047, AAA1682048, AAA1682049, AAA1682051, AAA1682052, AAA1682053, AAA1682054, AAA1692056, AAA1692057, AAA1692058, AAA1692059, AAA1692060, AAA1712011, AAA1712012, AAA1712013, AAA1712014, AAA1712015, AAA1712016, AAA1712017, AAA1722022, AAA1722023, AAA1722024, AAA1722036, AAA1722038, AAA1722039, AAA1722040, AAA1722041, AAA1722042, AAA1722043, AAA1722044, AAA1722045, AAA1732046, AAA1732047, AAA1732049.
510(k) numbers
["K151451"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register