FDA Medical Device Recalls (openFDA)
Urologix, Inc.: The following kits contain the UROLOGIX, Coolant Bag, CA1111B, Sterile EO, Manufactured for: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. UROLOGIX, CTC Advance Short, (2.5-3.5 cm), PT1034, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. CTC Advance Standard , (3.0-5.0 cm), PT1036, Sterile EO,, Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. CTC Advance Long, (4.5+ cm), PT1038, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. Targis Standard, (3.0-5.0 cm), PT1013, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. The Urologix Targis system is a non-surgical device intended to relieve symptoms and obstruction associated with Benign Prostatic Hyperplasia (BPH) and is indicated for men with prostatic urethral lengths greater than 2.5 cm.
- Recall number
- Z-2710-2011
- Recalling firm
- Urologix, Inc.
- Product
- The following kits contain the UROLOGIX, Coolant Bag, CA1111B, Sterile EO, Manufactured for: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. UROLOGIX, CTC Advance Short, (2.5-3.5 cm), PT1034, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. CTC Advance Standard , (3.0-5.0 cm), PT1036, Sterile EO,, Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. CTC Advance Long, (4.5+ cm), PT1038, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. Targis Standard, (3.0-5.0 cm), PT1013, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. The Urologix Targis system is a non-surgical device intended to relieve symptoms and obstruction associated with Benign Prostatic Hyperplasia (BPH) and is indicated for men with prostatic urethral lengths greater than 2.5 cm.
- Status
- Terminated
- Initiated
- 2011-04-28
- Posted
- 2011-06-30
- Terminated
- 2012-04-26
- Reason
- Urologix is recalling certain lots of Targis System Coolant Bag devices distributed over a two week period. The devices are packaged into disposable Procedure Kits containing a microwave catheter, a Coolant Bag and a RTU Plus balloon. The reason for this Recall involves three lots of Coolant Bags that contain an increased risk of leakage of water at the Sensor Module. This leak is detectabl
- Root cause
- Nonconforming Material/Component
- Action
- Urologix, Inc. sent an "IMPORTANT RECALL INFORMATION" letter dated April 28, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs the customers to identify the affected kit lots, remove the Coolant Bag pouch(s) and to complete and return a Return Material Authorization form with the product(s). Additionally, an enclosed Customer Acknowledgement Form should be completed and faxed to 763-475-1443. Customers with questions should contact the Director of Regulatory and Quality at 763-475-1400.
- Quantity
- 119
- Distribution
- Nationwide Distribution--including the states of AL, CA, CO, FL, GA, MA, MI, MN, NJ, NY, OR, PA, TN, WV, and including Puerto Rico.
- Lot, serial or model codes
- Catalog # / Lot No. PT1034 / AHP0333111A, AHP040411A, AHP040411B, AHP040611A, AHP040711A, AHP040811A, AHP041211A PT1036 / ACP033011B, ACP040111A, ACP040111B, ACP040611A, ACP040711A, ACP040711B, ACP040811A, ACP041211A PT1038 / ALP033111A, ALP040411A PT1013 / KP040711A CA1111B / C900284, C900285 and C900286 C900284, C900285 and C900286 C900284 and C900285 C900285
- 510(k) numbers
- not on file
- PMA numbers
- ["P970008"]
- Product code
- MEQ
- Firm FEI
- 3000718675
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28