Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Brilliance CT System 6/10/16/16P/40 configuration. No packaging, sold as single unit. Computed tomography scanner intended to provide cross sectional images of the human body and visualization of the internal organs.

Recall number
Z-2709-2014
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance CT System 6/10/16/16P/40 configuration. No packaging, sold as single unit. Computed tomography scanner intended to provide cross sectional images of the human body and visualization of the internal organs.
Status
Terminated
Initiated
2005-12-27
Posted
2014-09-26
Terminated
2015-11-06
Reason
The system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.
Root cause
Device Design
Action
Customers will be contacted by a field service technician with the field correction order. The firm will install the new labeling to the device.
Quantity
364
Distribution
Worldwide Distribution -- AZ, CO, GA, IA, IL, IN, KS, KY, MD, MI, MN, MS, NC, NY, OH, OK, TX, VT, and WI; and, countries of Singapore, New Zealand, India, Australia, Korea, Japan, Korea, China, India, Taiwan, Australia, Turkey, France, South Africa, Spain, Italy, Austria, Belgium, Switzerland, Israel, Netherlands, England, and Brazil.
Lot, serial or model codes
728256, 728251, 728246, 728240, and 728235.
510(k) numbers
["K010817","K033326"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register