Skip to the content

FDA Medical Device Recalls (openFDA)

Medacta Usa Inc: Product Brand Name: Versafitcup Double Mobility Liner Inserter Product Generic Name: Versafitcup Double Mobility Liner Inserter Model Number: 01.26.10.0018 The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature. The patients condition should be due to one or more of: " Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriactic arthritis, Congenital hip dysplasia, Ankylosing spondylitis. " Avascular necrosis of the femoral head. " Acute traumatic fracture of the femoral head or neck. " Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present. " Dislocation risks.

Recall number
Z-2709-2011
Recalling firm
Medacta Usa Inc
Product
Product Brand Name: Versafitcup Double Mobility Liner Inserter Product Generic Name: Versafitcup Double Mobility Liner Inserter Model Number: 01.26.10.0018 The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature. The patients condition should be due to one or more of: " Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriactic arthritis, Congenital hip dysplasia, Ankylosing spondylitis. " Avascular necrosis of the femoral head. " Acute traumatic fracture of the femoral head or neck. " Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present. " Dislocation risks.
Status
Terminated
Initiated
2011-05-23
Posted
2011-06-30
Terminated
2011-07-12
Reason
Medacta International has recently realized that the base of the Liner Inserter Ref 01.26.10.0018 Lot 105549 of the Versafitcup Double Mobility system could become rusty after the washing procedures. This is due to an incorrect raw material used by one of our suppliers during the production phase of this lot. For this reason, Medacta International has made the decision to remove all the pieces
Root cause
Labeling False and Misleading
Action
Medacta USA sent an "URGENT SAFETY NOTIFICATION-VERSAFITCUP DM LINER INSERTER" letter dated May 23, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to check their inventory for the affected device, immediately sequester it and not use the device in surgery. The letter instructs customers to pack the affected device for shipment to Medacta USA under Return Material Authorization 1590. Customer will receive a replacement within a few days. Questions regarding this recall are directed to the Director of Regulatory and Quality at 805-437-7085 ext. 26.
Quantity
Worldwide: 18 pieces. In the USA: 9 pieces
Distribution
Worldwide Distribution--USA (nationwide) including the states of CA, ID, IL, and OH, and 9 products to foreign countries.
Lot, serial or model codes
Lot code: 105549
510(k) numbers
["K083116"]
PMA numbers
not on file
Product code
MEH
Firm FEI
3006639916
Firm city
Camarillo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register