Staar Surgical Co.: Brand Name STAAR Surgical Collamer Ultraviolet-Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens Common Names Collamer single piece IOL Collamer plate haptic IOL The Device is intended to correct aphakia in persons 60 years of age or older in whom a cataractous lens has been removed by cataract extraction. The Device is to be implanted in the posterior chamber and in the capsular bag through a tear-free capsulorhexis (circular tear anterior capsulotomy).

Recall number
Z-2706-2011
Recalling firm
Staar Surgical Co.
Product
Brand Name STAAR Surgical Collamer Ultraviolet-Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens Common Names Collamer single piece IOL Collamer plate haptic IOL The Device is intended to correct aphakia in persons 60 years of age or older in whom a cataractous lens has been removed by cataract extraction. The Device is to be implanted in the posterior chamber and in the capsular bag through a tear-free capsulorhexis (circular tear anterior capsulotomy).
Status
Terminated
Initiated
2011-05-20
Posted
2011-06-29
Terminated
2014-03-04
Reason
The recall was initiated by Staar Surgical due to an error in the Directions for Use (DFU) for Staar Surgical's Collamer Ultraviolet Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens (Collamer IP IOL), also known as the nanoFLEX IOL. The Device Description erroneously indicates that the lens could be implanted in the ciliary sulcus.
Root cause
Other
Action
Staar Surgical send out an URGENT - Medical Device Correction Notice to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to acknowledge receipt of the letter by signing, dating, and faxing the form to Staar Surgical. For any questions call (800) 292-7902, ext 2345.
Quantity
16,750 devices distributed to the field
Distribution
Worldwide Distribution - USA (nationwide) and the countries of South Africa, Slovakia, Mexico, Hong Kong, and Sri Lanka
Lot, serial or model codes
CC4204BF and CC4204A
510(k) numbers
not on file
PMA numbers
["P990013"]
Product code
HQL
Firm FEI
2023826
Firm city
Monrovia
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register