Siemens Healthcare Diagnostics, Inc.: Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.

Recall number
Z-2703-2026
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.
Status
Open, Classified
Initiated
2026-06-08
Posted
2026-07-08
Terminated
not on file
Reason
The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.
Root cause
Under Investigation by firm
Action
Siemens Healthineers notified consignees on about 06/08/2026 via letter. Consignees were instructed to perform QC on each reagent well used for all A1c_E/A1c_H testing and enable "QC on Pack Change By Assay Type" to ensure QC is performed whenever testing transitions between wells. Enabling this setting will apply QC on Pack Change for all CH assays. If the lab uses Atellica CH to monitor QC, enable Patient QC Flagging to allow identification of potentially impacted patient results following a QC failure. Potentially impacted patient results will display a QC Fail flag and must be held from reporting and rerun after passing QC is obtained. If the lab uses Atellica Data Manager or other middleware to monitor QC, following a QC failure, hold all A1c_E/A1c_H samples processed after the failed QC event for review. Unload the reagent pack and load a new pack, repeat QC testing, and after acceptable QC results have been obtained, repeat any held A1c_E/A1c_H samples since the failed QC event. If repeated QC failures occur across consecutive wells, follow the lab s standard troubleshooting procedures to evaluate potential alternative causes for QC failure. Ensure to review the letter with the Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Consignees were requested to complete and return the Field Correction Effectiveness Check Form, retain the letter with laboratory records and forward this letter to those who may have received this product, and as a result of these actions, track reagent inventory, associated with implementation of these actions, to report to Siemens Healthineers once a resolution is provided. A follow-up communication will be provided once the Customer Actions described above are no longer required. The follow-up communication will include an option for reporting A1c_E reagent inventory, including additional and/or discarded A1c_E/A1c_H tests or wells, to Sieme
Quantity
24,077 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bangladesh, Bolivia, Brazil, Bulgaria, Burkina¿Faso, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Democratic Republic Congo, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Poland, Portugal, Qatar, Republic of Korea, Romania, Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Switzerland, Thailand, Turkey, Turkmenistan, U.A.E., Uganda, United¿Kingdom, Uruguay, Vatican City, Vietnam.
Lot, serial or model codes
Material Number: 11097536. UDI-DI: 00630414220505. All Lot Numbers.
510(k) numbers
["K200256"]
PMA numbers
not on file
Product code
PDJ
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register