Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated for use in surgical arthroscopic rotator cuff repair procedures. REF/ Catalog Number : 014771

Recall number
Z-2702-2020
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated for use in surgical arthroscopic rotator cuff repair procedures. REF/ Catalog Number : 014771
Status
Open, Classified
Initiated
2020-06-24
Posted
2020-07-24
Terminated
not on file
Reason
The T-Fix sterile package contains only one sleeve instead of two, The device will not perform as intended for the second bone tunnel due to the absence of thesleeve
Root cause
Process design
Action
Smith & Nephew issued Urgent Medical Device Recall Notice R-2020-12 dated June 24, 2020 via email and FedEx. to distributors, sales reps, hospitals or surgeons. Letter states reason for recall, health risk and action to take: Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com. Expanded Recall: An updated FSN was sent on August 28, 2020 to customers that received the additional affected batch. The recall was expanded as a result of additional complaints.
Quantity
137 Lot 2042492; 130 Lot 2049408
Distribution
Worldwide distribution - US Nationwide distribution including in the state of CO and the countries of AT, BE, CH, CL, DE, DK, ES, FR, GB, IT. Expanded 8/28/20 recall: 2nd Field Action: AU, CH, DE, FR, GB, IT, NL, PL, PT, & ZA
Lot, serial or model codes
Lot Number: 2042492 8/28/20 Expanded Recall- Lot number 2049408
510(k) numbers
["K925573"]
PMA numbers
not on file
Product code
HRX
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register