Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet U.K., Ltd.: Oxford Femoral Knee System REF 154600, LOT 2272736, WARNING DO NOT RESTERILIZE, STERILE UNLESS PACK DAMAGED, BIOMET UK Ltd, (510K# K011138) This device is used in partial knee replacement surgery as a result of patients who are experiencing degenerative arthritis.

Recall number
Z-2698-2011
Recalling firm
Biomet U.K., Ltd.
Product
Oxford Femoral Knee System REF 154600, LOT 2272736, WARNING DO NOT RESTERILIZE, STERILE UNLESS PACK DAMAGED, BIOMET UK Ltd, (510K# K011138) This device is used in partial knee replacement surgery as a result of patients who are experiencing degenerative arthritis.
Status
Terminated
Initiated
2011-04-22
Posted
2011-06-29
Terminated
2011-08-03
Reason
This recall was initiated due to a customer complaint which reported that while peeling off the Tyvek lid on the blister pack, very little effort was required to remove it and this could lead to contamination of the device.
Root cause
Packaging
Action
Biomet sent a "FIELD SAFETY NOTICE" dated April 21, 2011, to all affected accounts. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product, locate and remove from inventory and quarantine pending return to their Bioment distributor. Customers were also instructed to complete and return the attached "Fax Back Response Form" to their local Biomet distributor. For questions customers should contact their local Biomet Contact.
Quantity
12 devices
Distribution
Outside the US to Austria, Belgium, and Germany
Lot, serial or model codes
REF 154600, LOT 2272736
510(k) numbers
["K011138"]
PMA numbers
["P010014"]
Product code
HRY
Firm FEI
3002806535
Firm city
South Wales
Firm state
Bridgend
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register