FDA Medical Device Recalls (openFDA)

Eos Imaging: Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS software versions: 1.2.3.3788, 1.2.4.4286, 1.3.2.3373, 1.3.3.3671, 1.3.4.3740, 1.3.5.3915, 1.3.6.4272,1.4.1.4813,1.4.2.4850. EOS system is a digital radiography system used in general radiographic examinations. sterEOS workstation is a medical image review station for acceptance, transfer, display, storage, and digital processing of 2D X-ray images.

Recall number
Z-2692-2011
Recalling firm
Eos Imaging
Product
Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS software versions: 1.2.3.3788, 1.2.4.4286, 1.3.2.3373, 1.3.3.3671, 1.3.4.3740, 1.3.5.3915, 1.3.6.4272,1.4.1.4813,1.4.2.4850. EOS system is a digital radiography system used in general radiographic examinations. sterEOS workstation is a medical image review station for acceptance, transfer, display, storage, and digital processing of 2D X-ray images.
Status
Terminated
Initiated
2011-04-13
Posted
2011-06-28
Terminated
2012-01-18
Reason
Print true size, print requests from the EOS acquisition station or the sterEOS image review station may, in certain cases, be affected by a different enlargement factor during printing. T
Root cause
Software design
Action
EOS imaging sent a "FIELD SAFETY NOTICE" letter dated April 13, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to wait for the printing queue to empty before sending a second "print true queue" request with different print settings. Customers should then click the button shown on the instruction sheet to view the print queue. EOS imaging will provide a software update to correct all affected systems at no cost to the customers. An EOS representative will contact customers to schedule a time to upgrade the software. Questions can be answered by calling 33 1 55 25 63 90.
Quantity
5 EOS Systems, 10 sterEOS workstations
Distribution
Worldwide Distribution--USA (nationwide) including the states of CA, IA, MO, VA, and WI, and countries of Canada and Europe.
Lot, serial or model codes
Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS software versions: 1.2.3.3788, 1.2.4.4286, 1.3.2.3373, 1.3.3.3671, 1.3.4.3740, 1.3.5.3915, 1.3.6.4272,1.4.1.4813,1.4.2.4850.
510(k) numbers
["K071546","K101398"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3006165260
Firm city
Paris
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register