Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Endoscopy: Disposable Hemostasis Clip, DCH-7-230, Rx only, Single Use only, Cook Medical Product Usage: This device is used for endoscopic clip placement within the gastrointestinal tract for the purpose of endoscopic marking, hemostasis for mucosal sub-mucosal defects less thank 3cm in the upper GI tract, bleeding ulcers, arteries les than 2 mm, and polyps less than 1.5cm in diameter in the GI tract,. This device is not intended for the repair of GI tract luminal perforations.

Recall number
Z-2691-2011
Recalling firm
Cook Endoscopy
Product
Disposable Hemostasis Clip, DCH-7-230, Rx only, Single Use only, Cook Medical Product Usage: This device is used for endoscopic clip placement within the gastrointestinal tract for the purpose of endoscopic marking, hemostasis for mucosal sub-mucosal defects less thank 3cm in the upper GI tract, bleeding ulcers, arteries les than 2 mm, and polyps less than 1.5cm in diameter in the GI tract,. This device is not intended for the repair of GI tract luminal perforations.
Status
Terminated
Initiated
2011-06-07
Posted
2011-06-28
Terminated
2012-05-15
Reason
The disposable clip may not deploy after being positioned inside the patient.
Root cause
Device Design
Action
Wilson-Cook Medical, Inc. (Cook Endoscopy) sent an "URGENT: PRODUCT RECALL" notice dated June 7, 2011 and a "PRODUCT RECALL EFFECTIVENESS SURVEY' dated June 21, 2011 to all affected consignees. The notice described the product, problem, and actions to be taken. Medical users were instructed to quarantine any affected products and return to Cook Medical. The distributors were instructed to contact their customers and coordinate the return of the affected products to Cook Medical. The recalled notice and survey should be completed and returned to Cook Medical via fax to 812-339-7316 or scanned and emailed to CustomerRelationsNA@cookmedical.com. If you have any questions regarding this matter, please contact the Customer Relations Department at toll free-1-800-457-4500, press 4 then enter extension 2146 or by email at CustomerRelationsNA@cookmedical.com .
Quantity
307 units
Distribution
(USA) Nationwide Distribution including the states of: CA, HI, MA, ME, NC and NY.
Lot, serial or model codes
Lot numbers: W2965637, W2976210, W2978497, W2983595, W2986018, W2987394, W2987969, W2990400, and W2998860.
510(k) numbers
["K023903"]
PMA numbers
not on file
Product code
MND
Firm FEI
1037905
Firm city
Winston Salem
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register