FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Coulter PrepPlus 2, Part Number 378600, Product Usage: The COULTER PrepPlus 2, when used in combination with the COULTER TQPrep Workstation, is intended to prepare human whole blood for In Vitro Diagnostic (IVD) Use with cleared Beckman Coulter IVD applications on cleared Beckman Coulter flow cytometers (FC500, Navios and Navios EX). The COULTER PrepPlus 2 is a microprocessor-controlled pipetting and diluting system, designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.

Recall number
Z-2690-2017
Recalling firm
Beckman Coulter Inc.
Product
Coulter PrepPlus 2, Part Number 378600, Product Usage: The COULTER PrepPlus 2, when used in combination with the COULTER TQPrep Workstation, is intended to prepare human whole blood for In Vitro Diagnostic (IVD) Use with cleared Beckman Coulter IVD applications on cleared Beckman Coulter flow cytometers (FC500, Navios and Navios EX). The COULTER PrepPlus 2 is a microprocessor-controlled pipetting and diluting system, designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.
Status
Terminated
Initiated
2017-05-22
Posted
not on file
Terminated
2021-03-31
Reason
When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the absolute counts on only the first tube in a run.
Root cause
Software design
Action
The recalling firm initiated their recall by letter disseminated via email on 05/22/2017. The letter requested the following actions be taken: 1. Insert an empty daughter tube in place of a patient sample as the first tube in the carousel that is programmed for the addition of Flow-Count Fluorospheres. 2. Refer to the COULTER PrepPlus Operators Guide / PrepPlus 2 Instructions for Use (IFU) for instructions on adding Flow-Count Fluorospheres to daughter tubes. 3. Review all results following the flow cytometer data review instructions referenced in the associated IFU. Beckman Coulter also requested the consignee share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. If the consignee forwarded any of the affected product(s) listed above to another laboratory, They were asked to provide them a copy of the letter.
Quantity
473 units
Distribution
Worldwide Distribution
Lot, serial or model codes
All serial numbers currently in the field
510(k) numbers
["K130253"]
PMA numbers
not on file
Product code
PER
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register