Skip to the content

FDA Medical Device Recalls (openFDA)

Mindray DS USA, Inc. dba Mindray North America: A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. Part numbers 0632F-PA0000X (US domestic) and 0632B-00014 (international)

Recall number
Z-2689-2016
Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Product
A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. Part numbers 0632F-PA0000X (US domestic) and 0632B-00014 (international)
Status
Terminated
Initiated
2016-06-15
Posted
not on file
Terminated
2017-03-28
Reason
Potential for a leak to occur on the back-up O2 and air e-size cylinder yokes on the A7 Anesthesia Delivery System.
Root cause
Nonconforming Material/Component
Action
Mindray sent via certified mail with return receipt a recall letter dated June 15, 2016 to their affected customers. Mindray will replace the gasket on all e-size cylinder yokes on the affected A7 systems. Customer was instructed to contact their local Mindray Service Representative to arrange for this replacement. Customers can continue to use their A7 system while awaiting the replacement of the gasket(s). Customers can contact Ms. Diane Arpino, Director, Quality Operations and Regulatory Affairs via email to: d.arpino@mindray.com or via telephone to: (201)995-8407
Quantity
167 units (165 units - US) and (2 units - International)
Distribution
US Nationwide Distribution to AL, MA, MD, MN, MO, NE, NJ, OK, OR, PA and VA; and Canada
Lot, serial or model codes
P/N 0632F-PA0000X - (Domestic) and 0632B-PA00014 - (International)   Serial numbers FR-5B000341, FR-5B000342, FR-5B000343, FR-5B000348, FR-5B000349, FR-51000148, FR-51000149, FR-51000150, FR-51000151, FR-51000152, FR-51000153, FR-51000154, FR-51000159 , FR-51000161, FR-51000162, FR-51000163, FR-51000164, FR-51000165, FR-51000166, FR-51000167, FR-51000168, FR-51000170, FR-51000171, FR-52000172, FR-52000173, FR-52000175, FR-52000176, FR-52000177, FR-52000178, FR-52000180, FR-52000181, FR-52000182, FR-52000183, FR-52000184, FR-52000185, FR-52000186, FR-53000187, FR-53000188, FR-53000189, FR-53000190, FR-53000191, FR-53000192, FR-53000193, FR-53000194, FR-53000197, FR-53000198, FR-53000199, FR-53000200, FR-53000203, FR-53000205, FR-53000206, FR-53000207, FR-53000209, FR-53000211, FR-53000212, FR-53000213, FR-53000214, FR-53000215, FR-53000216, FR-53000217, FR-54000218, FR-54000220, FR-54000221, FR-54000223, FR-54000224, FR-54000228, FR-54000229, FR-54000231, FR-54000233, FR-54000234, FR-54000235, FR-54000236, FR-54000237, FR-54000238, FR-54000239, FR-54000240, FR-54000241, FR-54000242, FR-54000243, FR-54000244, FR-56000245, FR-56000246, FR-56000247, FR-56000248, FR-56000249, FR-56000250, FR-56000251, FR-56000252, FR-56000253, FR-56000254, FR-56000255, FR-56000256, FR-56000257, FR-56000258, FR-56000259, FR-56000260, FR-56000261, FR-56000262, FR-56000265, FR-56000266, FR-56000267, FR-56000268, FR-57000269, FR-57000270, FR-57000271, FR-57000272, FR-57000274, FR-57000275, FR-57000276, FR-57000277, FR-57000278, FR-57000279, FR-57000280, FR-57000281, FR-57000282, FR-57000283, FR-57000284, FR-57000285, FR-57000286, FR-57000287, FR-57000288, FR-58000289, FR-58000290, FR-58000291, FR-58000292, FR-58000293, FR-58000294, FR-58000295, FR-58000296, FR-58000297, FR-58000298, FR-58000299, FR-58000300, FR-58000301, FR-58000302, FR-58000303, FR-58000304, FR-58000305, FR-58000306, FR-58000307, FR-58000308, FR-58000312, FR-58000313, FR-58000314, FR-58000315, FR-58000316, FR-58000317, FR-58000318, FR-58000319, FR-58000320, FR-58000321, FR-58000322, FR-58000323, FR-58000325, FR-59000330, FR-59000331, FR-59000333, FR-59000334, FR-59000335, FR-59000336, FR-59000338, and FR-59000339.
510(k) numbers
["K151954"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register