Intrinsic Therapeutics, Inc.: Barricaid Annular Closure Impactor, REF: BAR-IMPACTOR, P/N: 400906-A

Recall number
Z-2688-2024
Recalling firm
Intrinsic Therapeutics, Inc.
Product
Barricaid Annular Closure Impactor, REF: BAR-IMPACTOR, P/N: 400906-A
Status
Open, Classified
Initiated
2024-07-25
Posted
2024-08-20
Terminated
not on file
Reason
At least two units have been found to be missing a weld.
Root cause
Process control
Action
On July 25, 2024 URGENT MEDICAL DEVICE RECALL letters were sent to consignees. Consignee Action Requested: Consignees are requested to take the following steps: 1. Prior to surgery, identify if any Impactors from the affected lots are in your inventory. Prioritize use of Impactors from lots not listed in this Recall letter and ensure that you bring at least two Impactors to the case. 2. Prior to surgery, share the contents of this Recall letter with the surgeon, highlighting the manufacturing defect that has been found and the potential clinical implications. 3. In most cases, there is no need for the Impactor when adhering to the surgical technique manual for Barricaid implantation. While proctoring, remind the surgeon to keep the endplate guide along the surface of the endplate of the vertebral body to be implanted and ensure the distal end of the delivery instrument is against the posterior wall of the vertebra. Confirm proper position before and during malleting by using fluoroscopy. 4. Ensure that If using an Impactor from the affected lot, the surgeon is aware that the tool shaft may be compromised. Use the Impactor per the surgeon technique manual and with fluoroscopic guidance for initial alignment of the instrument and after every 1-2 mallet strikes. If any bending at the weld location is observed, discontinue use of that Impactor and open a second instrument from inventory. 5. If no additional Impactors are available and the Barricaid implant is proud, remove the Implant using a Barricaid Removal Tool. Do not allow the surgeon to attempt to advance the implant further through use of any other instrument as there is a risk of damaging the implant itself. 6. Alternate Impactors will be shipped to you with instructions on returning any Impactors that you may have from the affected lots. Intrinsic will be in touch to provide product return instructions and replace your inventory. 7. Expeditiously fill out the attached acknowledgement form. The completed form c
Quantity
207 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
REF: BAR-IMPACTOR, P/N: 400906-A; UDI-DI: M906IMP1A0; Lot numbers: 10282203, 06152307, 08292301, 10112307.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3006232063
Firm city
Woburn
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register