Skip to the content

FDA Medical Device Recalls (openFDA)

Preat Corp: Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital Analog, REF 9001303 3i Certain-compatible 5.0mm Digital Analog, REF 9001903 3i Certain-compatible 6.0mm Digital Analog, REF 9002003 Astra-compatible 3.0mm Digital Analog, REF 9002103 Astra-compatible Aqua 3.5/4.0mm Digital Analog, REF 9002203 Astra-compatible Lilac 4.5/5.0mm Digital Analog, REF 9002203 BioHorizons-compatible 3.0mm Digital Analog, REF 9005203 Legacy-compatible 3.0mm Digital Analog, REF 9005303 Digital Analog - BioHorizons - 3.5mm, Digital Analog - BioHorizons - 4.5mnm, REF 9005403 Legacy-compatible 3.5mm Digital Analog, REF 9005503 Legacy-compatible 5.7mm Digital Analog, REF 9008503 Astra EV 4.8mm Milled Titanium Abutment REF 9006767 Digital analogs are used for visual inspection of abutments and restoration in the dental laboratory.

Recall number
Z-2687-2024
Recalling firm
Preat Corp
Product
Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital Analog, REF 9001303 3i Certain-compatible 5.0mm Digital Analog, REF 9001903 3i Certain-compatible 6.0mm Digital Analog, REF 9002003 Astra-compatible 3.0mm Digital Analog, REF 9002103 Astra-compatible Aqua 3.5/4.0mm Digital Analog, REF 9002203 Astra-compatible Lilac 4.5/5.0mm Digital Analog, REF 9002203 BioHorizons-compatible 3.0mm Digital Analog, REF 9005203 Legacy-compatible 3.0mm Digital Analog, REF 9005303 Digital Analog - BioHorizons - 3.5mm, Digital Analog - BioHorizons - 4.5mnm, REF 9005403 Legacy-compatible 3.5mm Digital Analog, REF 9005503 Legacy-compatible 5.7mm Digital Analog, REF 9008503 Astra EV 4.8mm Milled Titanium Abutment REF 9006767 Digital analogs are used for visual inspection of abutments and restoration in the dental laboratory.
Status
Open, Classified
Initiated
2024-06-25
Posted
2024-08-19
Terminated
not on file
Reason
Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.
Root cause
Process control
Action
On 07/22/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter via certified mail informing customers that Digital Analogs were manufactured out of specification for the rotational feature which may result in the dental restoration/implant may be slightly different than the position of the actual implant in the patient's mouth. Customers are instructed to: Product in the scope of this recall should be discarded or returned to PREAT CORPORATION. For questions contact Quality Regulatory Affairs at 224-622-7191 or email jespino@younginnovations.com
Quantity
185 devices
Distribution
U.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI, MN, NY, OH, OK, OR, PA, RI, TX, UT, and WA
Lot, serial or model codes
Model Numbers: 9000703, 9001303, 9001903, 9002003, 9002103, 9002203, 9002203, 9005203, 9005203, 9005303, 9005403, 9005503, 9008503, 9006767 UDI-DI codes: 00842092178690, 00842092181898, 00842092181959, 00842092178737, 00842092178775, 00842092178812, 00842092178812, 00842092178898, 00842092178898, 00842092178997, 00842092178928, 00842092178959, 00842092182130 Lot Numbers: 250219, 248890, 251432, 250602, 250418, 250213, 251459, 249071, 250335, 250223, 250336, 249510, 250215.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NDP
Firm FEI
2918719
Firm city
Santa Maria
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register